RecallDepth

CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-51; 3) PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA, Model Number: 21-2112-0100-51; 4) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2112-0300-01; 5) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2111-0300-01; 6) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-51; 7) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 8) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-51; 9) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 10) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, CE ENGLISH 1/EA, Model Number: 21-2111-0400-51; 11) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2111-0300-50; 12) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 13) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 14) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-78; 15) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 16) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2112-0300-00; 17) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-78; 18) PUMP, DEMO, CADD-SOLIS, MODEL 2100, NOT FOR HUMAN USE 1/EA, Model Number: 21-2101-249; 19) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-78; 20) PUMP KIT, CADD-SOLIS, YELLOW, MDL 2100 1/EA, Model Number: 21-2102-51; 21) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, NFHU, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-249; 22) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA, Model Number: 21-2111-0300-231; 23) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-78; 24) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0400-51; 25) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2112-0300-50; 26) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 27) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 28) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 29) PUMP, CADD-SOLIS, MODEL 2110, CE ENGLISH OUS 1/EA, Model Number: 21-2111-0100-50; 30) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 31) PUMP KIT, CADD-SOLIS HPCA, YELLOW KEYPAD, MODEL 2110, CE ENGLISH, OUS 1/EA, Model Number: 21-2112-0100-50; 32) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 33) PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA, Model Number: 21-2112-0400-01.

Company
Smiths Medical Asd Inc.
Recall Initiated
April 10, 2025
Posted
May 5, 2025
Recall Number
Z-1681-2025
Quantity
313,676 total units
Firm Location
Minneapolis, MN

Reason for Recall

There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

Distribution

Worldwide - US Nationwide distribution.

Lot / Code Info

All serial numbers. 1 ) 21-2111-0402-51 , UDI-DI: 15019517154290 ; 2 ) 21-2112-0402-51 , UDI-DI: 15019517154993 ; 3 ) 21-2112-0100-51 , UDI-DI: 10610586039232 ; 7 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 9 ) 21-2101-51 , UDI-DI: 10610586031854 ; 13 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 14 ) 21-2111-0402-78 , UDI-DI: 15019517154887 ; 15 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 16 ) 21-2112-0300-00 , UDI-DI: 10610586038822 ; 17 ) 21-2112-0402-78 , UDI-DI: 15019517155013 ; 18 ) 21-2101-249 , UDI-DI: 10610586034787 ; 20 ) 21-2102-51 , UDI-DI: 10610586033933 ; 22 ) 21-2111-0300-231 , UDI-DI: 10610586038785 ; 23 ) 21-2112-0401-78 , UDI-DI: 15019517101034 ; 24 ) 21-2112-0400-51 , UDI-DI: 10610586041051 ; 25 ) 21-2112-0300-50 , UDI-DI: 10610586038860 ; 26 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 27 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 28 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 29 ) 21-2111-0100-50 , UDI-DI: 10610586039256 ; 30 ) 21-2101-51 , UDI-DI: 10610586031854 ; 31 ) 21-2112-0100-50 , UDI-DI: 10610586039270 ; 32 ) 21-2101-00 , UDI-DI: 35019517154904; 33 ) 21-2112-0400-01 , UDI-DI: 10610586041822 ;

Root Cause

Under Investigation by firm

Action Taken

An URGENT MEDICAL DEVICE CORRECTION dated 4/10/25 was sent to customers. Actions for Users: Inform all affected CADD-Solis and CADD-Solis VIP users of this notice. Provide the instructions below: 1. Ensure all users or potential users of these products are immediately made aware of this notification. 2. Upon receipt of this notification and ongoing during regular maintenance, visually inspect the external condition of the battery pack or Wireless Communication Module and look for evidence of any physical damage. Handle a damaged device with caution. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 3. Visually inspect the condition of the battery compartment for evidence of any physical damage. Contact Technical Support for repairs. 4. Always check the battery compartment for fluid or debris before inserting the batteries, and do not allow any fluid or debris to fall into the battery compartment. 5. Complete the electronic Response Form via the link provided within this email notification within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product. 6. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward your customer list to Sedgwick at icumedical3393@sedgwick.com. Sedgwick will notify customers with email addresses via email and will send physical notifications to customers without email addresses in order for them to complete the Response Form Follow up Actions by Smiths Medical: Smiths Medical is continuing to investigate this matter to determine if additional actions may be warranted. For further inquiries, please contact Smiths Medical using the following information: Smit

More recalls by Smiths Medical Asd Inc.

View all recalls by this company →