Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z
- Company
- Maquet Cardiovascular, Llc
- Recall Initiated
- September 13, 2023
- Posted
- October 22, 2023
- Recall Number
- Z-0137-2024
- Quantity
- 2951 units
- Firm Location
- Wayne, NJ
Reason for Recall
Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.
Distribution
Nationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR
Lot / Code Info
UDI-DI: 00607567100008 Lot Numbers: 3000188485, 3000193165, 3000193670, 3000195673, 3000198652, 3000199944, 3000201253, 3000204442
Root Cause
Device Design
Action Taken
Maquet /Getinge issued Urgent Device Medical Device Removal (FSCA 2242352-09/06/2023-004-R) via FedEx with delivery receipt on 9/13/23. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the Acrobat-i Stabilizer System with the product codes/lot numbers listed in this notice. If you have unused/unexpired affected product that you will be returning from your inventory, please contact Maquet/Getinge Customer Service at 888-880-2874 between the hours of 6 AM and 5 PM Pacific Standard Time to request a return authorization number (RMA) and shipping instructions. If you have affected product, you are entitled to a credit. You will receive credit upon your acknowledgement that you have affected product for return. Please complete and sign the attached MEDICAL DEVICE Removal - RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by emailing a scanned copy to acrobati-mountbreak2023.act@getinge.com or by faxing the form to (877) 634-6230. Please forward this information to all current and potential Acrobat-i Stabilizer System users within your hospital / facility.