Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T

Company
Origen Biomedical, Inc.
Recall Initiated
December 19, 2023
Posted
February 9, 2024
Recall Number
Z-1068-2024
Quantity
3613 Cases
Firm Location
Austin, TX

Reason for Recall

Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.

Distribution

US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE

Lot / Code Info

REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281

Root Cause

Nonconforming Material/Component

Action Taken

On 11/19/23, recall notices were distributed to customers who were asked to do the following: 1) Identify and quarantine affected devices. 2) Further distribution or use should cease immediately. 3) Return affected devices to the recalling firm. 4) Complete and return the customer reply form via email to info@origen.com Customers with questions can contact the recalling firm at 1-800-233-9014 between the hours of 08:00 and 16:30 (CST). On 1/3/24, the firm sent out an additional recall communication removing one of the affected lot numbers.