ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV13F

Company
Origen Biomedical, Inc.
Recall Initiated
January 17, 2017
Posted
June 4, 2019
Terminated
July 27, 2021
Recall Number
Z-1230-2019
Quantity
145
Firm Location
Austin, TX

Reason for Recall

Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.

Distribution

US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile

Lot / Code Info

Lots: N18573, N18687-1

Root Cause

Process control

Action Taken

On 01/17/2017, Technical Bulletin 17.01 was mailed to customers. Customers were informed to clamp all sizes of catheters only in the middle. The following contact information was provided: US: 1-512-474-7278, and Germany: +49-7475-915591.