RecallDepth

Samsung Digital Diagnostic Mobile X-ray System, Model GM85.

This recall is currently active, issued February 9, 2024. It was issued by Neurologica Corporation.

Company
Neurologica Corporation
Recall Initiated
December 28, 2023
Posted
February 9, 2024
Recall Number
Z-1056-2024
Quantity
261 devices
Firm Location
Danvers, MA
Official Source
View on FDA website ↗

Reason for Recall

Mobile x-ray systems moving arm frame welding issue which presents the potential of bodily harm due to fall of the arm.

Distribution

US Nationwide Distribution including states of: AL, AR, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SD, TN, TX, VA, WI & WV.

Lot / Code Info

UDI-DI: 08806088582412, serial numbers: 5143M3HK400010M, 5143M3IJ800012D, 5143M3IK400005T, 5149M3IK400001H, 5149M3IK400002K, 5143M3IK400003Z, 5143M3GK500001H, 513YM3IK100001A, 5143M3HJ500001Y, 5143M3IJ600014H, 5143M3HK400002V, 5143M3JK400005Y, 5143M3HJ900001L, 5143M3IJ400007P, 5143M3HK400007T, 5143M3HK400003P, 513YM3IJA00001X, 5143M3JJ400007E, 5143M3IJ400006V, 5143M3JJ400006K, 5143M3JK400009V, 5143M3IJ200001D, 5143M3JJC00004Y, 5143M3JJ200017Y, 5143M3IHC00009M, 5143M3IJ100011V, 5143M3JJ300001K, 5143M3HJ600003K, 5143M3IK300002J, 5143M3JJ300003J, 5143M3JJ300002X, 5143M3JK400011A, 5143M3HK300016Y, 5143M3IJ800008L, 5143M3IJ600013B, 5143M3IJ100010M, 5143M3IJ100002P, 5143M3HK300011R, 5143M3HK200004H, 5143M3HK200006K, 5143M3JJC00001X, 5143M3IJ100009B, 5143M3IJ100006J, 5143M3HK300017E, 5143M3JK400006R, 5143M3JK400007B, 5143M3JK400004M, 5143M3HK300004E, 5143M3IJ600011M, 5143M3JK100012J, 514GM3HK100005L, 5143M3JK100003Z, 5143M3JK100002B, 5143M3IJC00001Y, 5143M3IK400002J, 5143M3IK400001P, 5143M3HK300007N, 514GM3HK100002R, 5143M3HK300010P, 5143M3HJA00002K, 514GM3IK400001L, 5143M3JJ200011L, 5143M3JJ300012N, 5143M3RJ100001M, 5143M3IJ600007N, 5143M3IHC00015L, 5143M3IHC00013X, 5143M3IHC00004N, 5143M3IHC00003Y, 5143M3JK400008N, 5143M3JK100015B, 5143M3HJA00006P, 5143M3HJ900007R, 5143M3JJ200012Z, 5143M3IK400008A, 5143M3JK400013K, 5143M3JJ700004K, 5143M3IHC00012Z, 5143M3JJ700001V, 514GM3JK300001A, 5143M3IK300008R, 5143M3IJ100001Y, 5143M3IHC00019T, 5143M3HK300001F, 5143M3HJA00010W, 5143M3IHC00018J, 513YM3IK400001W, 5143M3IJ100003X, 5143M3HK400009W, 5143M3JJ500002M, 5143M3IHC00008E, 5143M3HK300019Z, 5143M3JJ200016T, 5143M3JK400010W, 5143M3IJ800013Y, 5143M3JK500001M, 5143M3JK500006X, 5143M3JK100006N, 5143M3IJ300004W, 5143M3JJ200014F, 5143M3HJ900003W, 5143M3HJ900002V, 5143M3IJ400003R, 5143M3IJ400004B, 5143M3JJ200005X, 5143M3JJ200004Y, 5143M3RJ100002V, 5143M3IHC00014H, 5143M3IJ600008K, 5143M3IJ200002F, 5143M3JJ200007W, 5143M3HJ900006H, 5143M3JJ300011Z, 5143M3JJ200002M, 5143M3JJ200001F, 5143M3HJA00005A, 5143M3HJA00004Y, 5143M3GK500002Y, 5143M3HJA00007D, 5143M3HJ900004M, 5143M3HJ900005J, 5143M3IJ600006W, 5143M3HJ500005P, 5143M3HJ500006W, 5143M3JJ400002X, 5143M3DHB00001M, 5143M3HK300018W, 514GM3HK100001E, 5143M3JJ700003D, 5143M3JJ700002R, 5143M3JK100014L, 5143M3HK300006Z, 5143M3IJ800001B, 5143M3IJ800003J, 5143M3IJ800002V, 5143M3IJ800005X, 5143M3IJ800004F, 5143M3JJ700005W, 5143M3IHC00005J, 5143M3JK500007L, 5143M3JK500002P, 5143M3JK500005V, 5143M3JK500003W, 5143M3IK400009H, 5143M3HJ600005D, 513YM3IK500003Z, 5143M3IK400007W, 513YM3IK500001K, 513YM3IK500004H, 5143M3IJ100012T, 5143M3HJ500004M, 5143M3IJ600004F, 5143M3IJ800011E, 5143M3IJ600005Y, 5149M3IK400003E, 5143M3IJ800007K, 5143M3IJ500001B, 5143M3IJ100004E, 5143M3HJ900009E, 5143M3IJ800014B, 5143M3IHC00002L, 5143M3IK300001X, 5143M3HK300003Y, 5143M3JK100008H, 5143M3JJ200015M, 5143M3JK400003D, 5143M3IK400006X, 5143M3IJ100005R, 5143M3IJ600001H, 5143M3JJ200018X, 5143M3HJA00008X, 5143M3JK100007F, 5143M3IJ800009M, 5143M3IJ400005N, 5143M3IJ200003M, 513YM3IK500005Y, 513YM3IK100002D, 513YM3IK400002A, 513YM3IK100003N, 5143M3IK300006T, 5143M3HK400004J, 5143M3IJB00003H, 5143M3IHC00001H, 5143M3JJ200009A, 5143M3HK200005V, 5143M3GK500003J, 5143M3IJ300002Y, 5143M3IJ300006N, 5143M3HJ900008F, 5143M3IHC00011V, 5143M3JJ400005H, 5143M3RJ100003T, 5143M3HJ900010Z, 5143M3JJC00002H, 5143M3HK300012L, 5143M3JJ200008B, 5143M3JJ400001T, 5143M3IHC00006T, 514GM3JJ900001R, 5143M3IJ600003L, 5143M3JJ200010J, 5143M3HJ500003A, 5143M3IJ600002P, 5143M3HJ500002J, 5143M3HJA00003Z, 5143M3IK300009L, 5143M3IJC00002N, 5143M3IK400010R, 5143M3IJ300003E, 5143M3HJA00001B, 5143M3HK400008X, 5143M3JK400012H, 5143M3HK300013A, 5143M3JK100001L, 5143M3IJ600012X, 513YM3IK500002R, 5143M3HK400006F, 5143M3JK400014E, 5143M3JJ300007T, 5143M3JJ300006H, 5143M3IK400004F, 5143M3JK100004A, 514GM3HK100003J, 5143M3JJC00003L, 5143M3JJB00001F, 5143M3HJA00009N, 5143M3IJ600010Z, 5143M3JK100013W, 5143M3IK500001F, 5143M3JJ300008P, 5143M3IHC00017N, 5143M3IHC00016Y, 5143M3JK500004D, 5143M3IJ600009J, 5143M3IJ100013K, 5143M3IK300005H, 5143M3IK300007P, 5143M3HK300009K, 5143M3JK100009M, 5143M3JJ300010W, 5143M3GK500006P, 5143M3JK100010E, 5143M3JK100011R, 5143M3JK100005D, 5143M3IHC00010R, 5143M3JJC00006J, 5143M3JJC00005N, 5143M3JJ300009R, 5143M3HK400005Z, 514GM3HK100004W, 5143M3HK300008M, 5143M3HK300005W, 5143M3JJ300004B, 5143M3JJ300005D, 5143M3JJ200006R, 5143M3IJ300005Z, 5143M3IHC00007A, 5143M3IJ600015P, 5143M3JJ200013D, 5143M3HK300002V, 5143M3HJ600004J.

Root Cause

Process design

Action Taken

An "Urgent - Medical Device Correction" notice was issued via email on December 28, 2023, cosigned by representatives of Samsung Electronics Co., Ltd. and NeuroLogica Corp. The notice describes the product, problem and actions to taken. The customers were instructed to continue to use the GM85 systems and Service Engineers will contact you to arrange for on-site correction. Share this notice with users within your organization or anywhere the devices may have been transferred. Customer Reply Form are to be completed and returned via email to fsca@neurologica.com. Contact your local representative with questions email: SupportBI@bostonimaging.com, 1-800-976-0980.