GM60A. Digital Diagnostic Mobile X-Ray System.

Company
Neurologica Corporation
Recall Initiated
July 8, 2024
Posted
August 22, 2024
Recall Number
Z-2715-2024
Quantity
31 units
Firm Location
Danvers, MA

Reason for Recall

Potential for the anti-fall system of the device arm to fail.

Distribution

US Nationwide distribution in the states of AR, CT, GA, IN, KS, MA, MD, MO, MS, NC, ND, NJ, NV, OH, PA, VA, WV & DC.

Lot / Code Info

UDI: 08806088511153/ 08806088088266; Serial No. G69812, G69821, G69824, G69986, G69987, G71422, G71425, G71431, G71434, G71795, G71798, G71802, G72052, G72053, G72264, G72279, G72296, G72301, G72310, G72313, G72319, G72320, G72481, G72484, G72595, G72601, G72853, G72875, G73179, G73184, G73542; Samsung Serial No. 50WFM9BG900003M, 50WFM9BG900004J, 50WFM9BG900005H, 50WFM9BGA00001T, 50WFM9BGA00002B, 50WFM9BH300001E, 50WFM9BH300002W, 50WFM9BH300003Z, 50WFM9BH300004N, 50WFM9BH400001X, 50WFM9BH400002W, 50WFM9BH400003A, 50WFM9BH500001H, 50WFM9BH500002Y, 50WFM9BH600001P, 50WFM9BH600002L, 50WFM9BH600003F, 50WFM9BH600004Y, 50WFM9BH600005W, 50WFM9BH600006N, 50WFM9BH600007K, 50WFM9BH600008J, 50WFM9BH700002B, 50WFM9BH700001P, 50WFM9BH700003M, 50WFM9BH700004E, 50WFM9BH800002E, 50WFM9BH800003D, 50WFM9BH900001X, 50WFM9BH900002D, 50WFM9BHB00001T.

Root Cause

Device Design

Action Taken

Consignees were emailed an "Urgent -- Medical Device Correction; GM60A Digital X-ray Imaging System" notification the week of 7/8/2024. The notification instructs consignees to share the recall notification with those in their organization that utilize the affected devices and forward the notification if devices have been transferred. Affected systems may continue to be used. Consignees are to check on-hand devices to confirm if they are affected by the field action and return the response form back indicating any devices that require correction by email to fcsa@neurologica.com within 10 business days. Samsung Electronics will send service engineers to affected consignees to correct devices free of charge. Service will be coordinated by Samsung Electronics and should be completed by October 2024. Consignees with any questions are to contact their local Samsung Electronics representative.

More recalls by Neurologica Corporation

View all recalls by this company →