RecallDepth

Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: CATH LAB PACK, Model Number DYNJ58351A

This recall is currently active, issued July 20, 2023. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
May 18, 2023
Posted
July 20, 2023
Recall Number
Z-2190-2023
Quantity
540 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.

Distribution

Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.

Lot / Code Info

all lots within expiry, distributed from Medline between December 2017 - May 2023: Model Number DYNJ58351A, UDI/DI (EA) 10888277739390, UDI/DI (CS) 40888277739391

Root Cause

Device Design

Action Taken

Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704

More recalls by Medline Industries, Lp - Northfield