RecallDepth

Medline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004A LAVH MINOR DYNJG901001B CRANIOTOMY DYNJ908819B KIT CRANIOTOMY (only units from SKU DYNJG001004A LAVH MINOR were distributed)

Company
Medline Industries, Lp - Northfield
Recall Initiated
July 23, 2025
Posted
August 28, 2025
Recall Number
Z-2479-2025
Quantity
4 ea
Firm Location
Northfield, IL

Reason for Recall

Kits contain SureStep Foley Tray Systems recalled by BD that include incorrect product information insert sheets. If a mislabeled tray is used, patient risk ranges from minor injury to injury that requires medical intervention, such as infection or local/systemic allergic reaction (anaphylaxis).

Distribution

Domestic distribution to FL and IL. No OUS distribution reported.

Lot / Code Info

Medline Kit SKU Description UDI (EA) UDI (Case) Kit Lot Number CDS984076J GYN ROBOTIC 10198459283147 40198459283148 25DBH902 DYNJG001004A LAVH MINOR 10195327677329 40195327677320 25ELA892 DYNJG901001B CRANIOTOMY 10198459164316 40198459164317 25DLB031 DYNJ908819B KIT CRANIOTOMY 10198459102677 40198459102678 25DDA909

Root Cause

Error in labeling

Action Taken

Firm notified customers via email and first-class mail of the action. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.