Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
This recall is currently active, issued August 19, 2025. It was issued by Olympus Corporation Of The Americas.
- Recall Initiated
- July 25, 2025
- Posted
- August 19, 2025
- Recall Number
- Z-2338-2025
- Quantity
- 155 units
- Firm Location
- Center Valley, PA
- Official Source
- View on FDA website ↗
Reason for Recall
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
Distribution
US Nationwide distribution.
Lot / Code Info
Model Number: WA91327U. Lot Code: ESG-410 (SEP) WA91327U. UDI-DI: 04042761087698. Serial Numbers: 101696, 101697, 101698, 101699, 101700, 101701, 101702, 101703, 101704, 101705, 101706, 101707, 101708, 101709, 101710, 101919, 101920, 101921, 101922, 101923, 101924, 101925, 101926, 101927, 101928, 101929, 101933, 101934, 101935, 101936, 101937, 101940, 101941, 101942, 101945, 101946, 101947, 102148, 102149, 102150, 102151, 102152, 102153, 102154, 102156, 102157, 102158, 102159, 102160, 102161, 102162, 102164, 102165, 102167, 102168, 102170, 102172, 102173, 102174, 102175, 102176, 102177, 102178, 102179, 102180, 102181, 102182, 102183, 102184, 102185, 102186, 102187, 102188, 102189, 102190, 102191, 102193, 102194, 102195, 102198, 102199, 102200, 102266, 102427, 101932, 102155, 102163, 102197, 102660, 102661, 102662, 102663, 102664, 102665, 102666, 102667, 102668, 102669, 102670, 102671, 102672, 102673, 102674, 102675, 102676, 102677, 102678, 102679, 102680, 102681, 102682, 102683, 102684, 102685, 102686, 102687, 102688, 102689, 102690, 102691, 102692, 102693, 102694, 102695, 102696, 102697, 102698, 102699, 102700, 102701, 102702, 102703, 102704, 102705, 102706, 102707, 102708, 102709, 102710, 102711, 102712, 102713, 102714, 102715, 102716, 102717, 102718, 102719, 102721, 102722, 102723, 102724, 102725, 103922, 103930.
Root Cause
Component change control
Action Taken
Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.