Olympus Single use Biopsy Valve. Model/Catalog Number: MAJ-210. 20 units per box. The MAJ-210 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent reflux of body fluids.
- Recall Initiated
- January 29, 2026
- Posted
- March 3, 2026
- Recall Number
- Z-1488-2026
- Quantity
- 95,882 boxes (20 per box)
- Firm Location
- Center Valley, PA
Reason for Recall
Potential for rubber fragment detachment during use.
Distribution
US Nationwide distribution.
Lot / Code Info
Model/Catalog Number: MAJ-210. UDI-DI: 14953170152433, 14953170452069. All Lot Numbers. 20 units per box.
Root Cause
Under Investigation by firm
Action Taken
Olympus notified consignees on about 01/29/2026 via letter titled URGENT: FIELD SAFETY NOTICE. Consignees were instructed to carefully read the content of the notification, ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, and that they may continue using the devices as per the notification and instructions for use. Olympus is highlighting the importance of strictly adhering to the Inspection of the Biopsy Valve and Inserting and Withdrawing the Endo-therapy accessories sections in the IFU related to the detachment of fragment from the slit part of rubber. Consignees were requested to acknowledge receipt of the notification via their portal and if product has been further distributed to notify those consignees.