Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.
- Recall Initiated
- February 15, 2024
- Posted
- February 14, 2025
- Recall Number
- Z-1171-2025
- Quantity
- 20 units
- Firm Location
- Minnetonka, MN
Reason for Recall
Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
Distribution
US distribution to states of: AL, TX, NV
Lot / Code Info
UDI-DI 00855913001794, Lot Number 2122
Root Cause
Device Design
Action Taken
Augustine Temperature Management contacted their consignees by visit beginning on 02/15/2024 to replace the affected cables. The firm collected email confirmation that the original cables had either been returned or destroyed. If you have any questions, contact the Product Manager at 952-465-3526 or email: