RecallDepth

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia

Company
Augustine Temperature Management, Llc
Recall Initiated
December 16, 2024
Posted
February 20, 2025
Recall Number
Z-1199-2025
Quantity
3 units
Firm Location
Minnetonka, MN

Reason for Recall

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Distribution

US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

Lot / Code Info

UDI-DI UDI-DI 00855913001848, Serial Numbers: 1304384, 1304619, 1304620

Root Cause

Device Design

Action Taken

Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.