RecallDepth

LAPAROSCOPY PACK, Model No. CALS41F, CALS41F-01, CALS41G

Company
American Contract Systems Inc
Recall Initiated
February 6, 2025
Posted
March 14, 2025
Recall Number
Z-1331-2025
Quantity
303 kits
Firm Location
Tiffin, OH

Reason for Recall

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Distribution

US Nationwide distribution in the state of OH.

Lot / Code Info

1. Model No. CALS41F UDI-DI 00191072215727 Tray Lots 962241 70-050280 70-050870 70-051556 2. Model No. CALS41F-01 UDI-DI 00191072215727 Tray Lot 70-052412 3. Model No. CALS41G UDI-DI 00191072233332 Tray Lot 70-053051

Root Cause

Process control

Action Taken

The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.

More recalls by American Contract Systems Inc

View all recalls by this company →