AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER CCKM67F; 3)HIP PROTHESIS PACK, MODEL/ITEM NUMBER HSTH03N; 4)DTC ANTERIOR HIP PACK (PS 209459), MODEL/ITEM NUMBER MIAH65B; 5)DR WHIDD0N TOTAL HIP CPP, MODEL/ITEM NUMBER MPWH01B; 6)KNEE ARTHROSCOPY, MODEL/ITEM NUMBER SPKA57T; 7)ANTERIOR HIP PACK, MODEL/ITEM NUMBER WVAH40H.

Company
American Contract Systems Inc
Recall Initiated
July 14, 2025
Posted
August 28, 2025
Recall Number
Z-2472-2025
Quantity
588 units
Firm Location
Tiffin, OH

Reason for Recall

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Distribution

US distribution to states of: IL and LA

Lot / Code Info

1) MODEL/ITEM NUMBER AMAH40N: UDI/DI 00191072235473, Kit Lot Numbers: 42-7886011, 42-7990911; 2)MODEL/ITEM NUMBER CCKM67F: UDI/DI 00191072233578, Kit Lot Numbers: 70-053862, 70-053457, 70-053042, 70-052715, 70-052358, 70-051850, 70-051243; 3)MODEL/ITEM NUMBER HSTH03N: UDI/DI 00191072189615, Kit Lot Numbers: 42-7115711; 4)MODEL/ITEM NUMBER MIAH65B: UDI/DI 00191072235411, Kit Lot Numbers: 42-8134311; 5)MODEL/ITEM NUMBER MPWH01B: UDI/DI 00191072233271, Kit Lot Numbers: 5436801; 6)MODEL/ITEM NUMBER SPKA57T: UDI/DI 00191072235848, Kit Lot Numbers: 12-7400011, 12-7381311, 12-7358211, 12-7082111, 12-7157111, 12-7309911; 7)MODEL/ITEM NUMBER WVAH40H: UDI/DI 00191072233004, Kit Lot Numbers: 42-7897411, 42-7966211.

Root Cause

Under Investigation by firm

Action Taken

Owens & Minor issued an "URGENT: MEDICAL DEVICE CORRECTION" notice to its consignees on 07/15/2025 via email. The notice explained the problem, potential risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of all affected components. " Add warning labels to affected procedure trays to ensure the components are discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you receive a product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness for your consignees. " Complete the attached Recall Response Form and return as soon as possible to GM-OMRA_RECALLS@owens-minor.com. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com. If you have any questions, contact the Manager, Regulatory Affairs at 470-280-4277 or email to: GM-OMRA_RECALLS@owens-minor.com.