Modular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 12 mm, Product Code: 880-601/12

Company
Waldemar Link Gmbh & Co. Kg (Mfg Site) Oststr. 4-10 Norderstedt Germany
Recall Initiated
February 14, 2025
Posted
March 17, 2025
Recall Number
Z-1373-2025
Quantity
52 units

Reason for Recall

A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.

Distribution

Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IN, KS, MO, NJ, NV, OH TX, VA, WI and the countries of Germany, Italy, Spain, Switzerland.

Lot / Code Info

Product Code: 880-601/12; UDI-DI: 04026575443932; Lot Numbers: 1943274, 1943276, 2049172, 2049173, 2049175.

Root Cause

Labeling Change Control

Action Taken

On February 21, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken by the Customer/User: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

More recalls by Waldemar Link Gmbh & Co. Kg (Mfg Site) Oststr. 4-10 Norderstedt Germany