Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.

Company
Waldemar Link Gmbh & Co. Kg (Mfg Site) Oststr. 4-10 Norderstedt Germany
Recall Initiated
November 10, 2025
Posted
December 9, 2025
Recall Number
Z-0890-2026
Quantity
18 units

Reason for Recall

The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.

Distribution

US distribution to states of: AL, GA, and TX.

Lot / Code Info

Item Number: 15-0028/08; UDI-DI: 04026575175222; Lot Number: 2535004;

Root Cause

Process control

Action Taken

On November 10, 2025, URGENT: MEDICAL DEVICE RECALL letters were emailed to distributors. Actions to be taken: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions.