Life2000 Ventilator PKG, Hospital, a) REF BT20007, b) REF BT-20-0007, ventilatory support for the care of individuals

Company
Baxter Healthcare Corporation
Recall Initiated
January 25, 2023
Posted
February 24, 2023
Recall Number
Z-1140-2023
Quantity
135 units
Firm Location
Deerfield, IL

Reason for Recall

There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.

Distribution

Nationwide

Lot / Code Info

a) REF BT20007, UDI 00887761978089, Serial Numbers: L191001-005, L191001-003, L191001-004, L191001-002, L191001-001, L191212-022, L191212-023, L191212-021, L191212-024, L191212-025, L200114-011, L200114-004, L200114-008, L1911019-015, L1911019-021, L1911019-023; b) REF BT-20-0007, UDI 00887761978089, Serial Numbers: L1911019-016, L1911019-020, L1911019-022, L2002006-032, L2003004-117, L2003004-085, L2003004-091, L2003004-121, L2002040-026, L2002040-039, L2002040-041, L2002011-033, L2002040-005, L2003004-145, L2003004-130, L2003004-128, L2003004-125, L2003004-123, L2003004-077, L2002040-040, L2002040-042, L2003004-131, L2003004-137, L2003004-144, L2003004-127, L2003004-133, L2003004-129, L2003004-115, L2003004-083, L2003004-087, L2002040-043, L2003004-079, L2003004-073, L2003004-098, L2003004-150, L2003004-181, L2003015-181, L2003015-189, L2003015-188, L2003015-180, L2003015-201, L2003015-205, L2003015-197, L2003015-203, L2003015-185, L2003015-186, L2003015-196, L2003015-198, L2003015-179, L2003015-202, L2003015-204, L2003015-178, L2003015-206, L2003015-309, L2003015-311, L2003015-349, L2003015-276, L2003015-275, L2004002-027, L2004002-015, L2003015-399, L2003015-395, L2003015-393, L2003015-265, L2003015-389, L2003015-391, L2003015-396, L2003015-397, L2003015-398, L2004002-020, L2004002-003, L2003015-384, L2003015-392, L2003015-277, L2003015-273, L2004002-262, L2004002-256, L2004002-263, L2004002-241, L2004002-174, L2004002-246, L2004002-265, L2004002-197, L2004002-170, 120210001615, 120190000718, 120190000711, 120190001264, 120190000088, 120190001269, 120190000078, 120190000089, 120210000884, 120190001266, 120190000092, 120190000730, 120190000075, 120190000706, 120190000737, 120190001275, 120190000728, 120190000093, 120190000707, 120210001613, 120210001608, 120190000077, 120190000091, 120190000074, 120210001614, 120190000714, L2005055-250, 120190000081, L2006020-252, 120190000079, L2005055-873, 120190000090, 120190000076, 120190000722, L2005055-898

Root Cause

Under Investigation by firm

Action Taken

Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its healthcare providers, wholesalers, distributor/reseller, and durable medical equipment providers on 01/25/2023 by USPS First Class Mail. The notice explained the issue and the hazard and requested the following actions be taken by adressee: "Please follow the daily checks and preventive maintenance requirements below to ensure the best oxygen delivery with the Life2000 ventilation system when used with a third-party oxygen concentrator. 1- Daily Checks a) Check the CombO2 hose and Breathe Pillow Interface for kinks, dirt, moisture, or damage and contact Baxter Customer Service at 800-426-4224, option 3, if replacement parts are needed. Baxter recommends replacing the Breathe Pillow Interface every 90 days and the CombO2 hose every six months. b) Do not modify, alter, or extend tubing, connectors, or interface. Additionally, when using the Life2000 system with an oxygen concentrator, only use Baxter s approved 6-foot, 20-foot or 50-foot CombO2 hose. c) Address alarms as described in the Life2000 Quick Reference Guide (PL-20-0044-D) pages 17-19. d) Make sure the CombO2hose is correctly attached to the Life2000 compressor and the oxygen concentrator, as shown in Attachment A, Figure 1. Make sure all connections are tight. e) Ensure the setting on the oxygen concentrator remains at the prescribed liter flow while connected to the Life2000 system as shown in the Life2000 Extended Range Configuration Quick Start Guide (APR186301). If the liter flow drops when you connect to the Life2000 system, take the following steps: - Stop using the Life2000 system and go back to your traditional oxygen cannula and concentrator. - Call Baxter s Clinical Support team at 800-397-9071 for troubleshooting. f) Please notify your health care team of any concerns or seek medical attention if you are exhibiting signs of respiratory distress (increased breathing rate, wheezing, bluish color around the mouth, lips or fingernails,

More recalls by Baxter Healthcare Corporation

View all recalls by this company →