RecallDepth

Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

Company
Baxter Healthcare Corporation
Recall Initiated
January 13, 2026
Posted
February 13, 2026
Recall Number
Z-1368-2026
Quantity
5
Firm Location
Deerfield, IL

Reason for Recall

Pumps were released without full testing being performed, including occlusion alarm testing.

Distribution

States: Ohio Countries: United States

Lot / Code Info

Lot Code: GTIN 00085412610900, Serial Numbers 3815663, 3823468, 3824743, 3825234, and 3825599

Root Cause

Employee error

Action Taken

Firm notified the single customer on 1/13/26 via phone call. Customer was asked to work with on-site service technician to coordinate return of affected product.

More recalls by Baxter Healthcare Corporation

View all recalls by this company →