RecallDepth

Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

This recall is currently active, issued February 13, 2026. It was issued by Baxter Healthcare Corporation.

Company
Baxter Healthcare Corporation
Recall Initiated
January 13, 2026
Posted
February 13, 2026
Recall Number
Z-1368-2026
Quantity
5
Firm Location
Deerfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Pumps were released without full testing being performed, including occlusion alarm testing.

Distribution

States: Ohio Countries: United States

Lot / Code Info

Lot Code: GTIN 00085412610900, Serial Numbers 3815663, 3823468, 3824743, 3825234, and 3825599

Root Cause

Employee error

Action Taken

Firm notified the single customer on 1/13/26 via phone call. Customer was asked to work with on-site service technician to coordinate return of affected product.

More recalls by Baxter Healthcare Corporation

View all recalls by this company →