RecallDepth

Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2 Catalog Number: 410700

Company
Meridian Bioscience Inc
Recall Initiated
December 15, 2022
Posted
January 20, 2023
Recall Number
Z-1008-2023
Quantity
92 kits
Firm Location
Cincinnati, OH

Reason for Recall

Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2

Distribution

US Nationwide distribution in the states of AZ, CA, IN, KS, KY, LA, MI, MN, NC, NY, OR, TX.

Lot / Code Info

UDI: N/A Lot Numbers 410700P019, 410700P020, 410700P021, 410700P022, 410700P023, 410700P024, 410700P027

Root Cause

Process control

Action Taken

Meridian Bioscience issued Urgent Medical Device Recall on 12/15/22 via email. The letter states reason for recall, health risk and action to take: REQUIRED ACTIONS: " Inspect the rotor, ring, and chamber walls of the Revogene instrument for signs of leakage (e.g., dried spots). Contact Meridian Bioscience Technical Service if any signs of leakage were detected. " If leakage is suspected, clean the Revogene, MOCK PIEs, and surrounding workspace using the instructions in the Revogene operator's manual. " Review current inventory and segregate any remaining stock. " Discontinue use of any remaining stock. " Contact Meridian Technical Support 800.343.3858 to obtain a FedEx label to return any remaining inventory to Meridian Bioscience Inc. for credit. " Complete and return the Customer Notification Form. This will indicate receipt of this field correction notice. Complete this form even if you have no remaining inventory. Questions, please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at MBI_TechService@meridianbioscience.com.

More recalls by Meridian Bioscience Inc

View all recalls by this company →