Compella Bariatric Bed System

Company
Baxter Healthcare Corporation
Recall Initiated
February 14, 2022
Recall Number
Z-0820-2022
Quantity
5186 units
Firm Location
Deerfield, IL

Reason for Recall

Potential damage to the power cord may cause the power conductors to be exposed, thereby creating a risk of shock from the power mains.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Cameroon, Canada, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Jamaica, Jordan, Kuwait, Mexico, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Romania, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkmenistan, UAE, and UK.

Lot / Code Info

All lots/serial numbers UDI information: Compella Bariatric Bed - 00887761000902

Root Cause

Labeling design

Action Taken

Customer notification letters were mailed beginning 02/21/2022. A localized customer letter will be mailed (with signature service) to all first-level consignees by Sedgwick. The letter will inform the customer of the issue, reason for Field Action and provide risk mitigations. The customer will be instructed to sign acknowledgement form to confirm receipt and return to a dedicated email address. Distributors will be asked to forward to their end users. Distributors will be provided with instructions on communication with their end users. All consignees will be required to return a signed acknowledgment form and confirm that they read and understand the issue. Distributors will be asked to forward to their end users. Distributors will be provided instruction on communication with their end users.

More recalls by Baxter Healthcare Corporation

View all recalls by this company →