EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. Model/Catalog Number: BD-410X-1055

Company
Olympus Corporation Of The Americas
Recall Initiated
September 13, 2024
Posted
October 25, 2024
Recall Number
Z-0225-2025
Quantity
209 units
Firm Location
Center Valley, PA

Reason for Recall

Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch

Distribution

Worldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA, TN, VA, WI; and Foreign OUS to countries of: CA, DE, IN, SG, Latin America.

Lot / Code Info

UDI-DI: 821925033238 Lot: 408987

Root Cause

Under Investigation by firm

Action Taken

Olympus issued Urgent Medical Device Removal letter issued 9/13/24 to Endoscopy Department, Risk Management Department. Letter states reason for recall, health risk and action to take: 2. Examine your inventory and identify the above-listed device with the affected batch number. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. When it is received, you will receive a credit for your affected device(s). 4. If you have further distributed this product, identify your customers, and forward them this notification. 5. Olympus requests that you acknowledge receipt of this letter. Access the Olympus recall portal to indicate that you have received this notification. a) Go to https://olympusamerica.com/recall b) Enter the recall number "0453 c) Complete the form as instructed. Olympus requests that you report any complaints, including any incorrect labeling, to the Olympus Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). If you require additional information, please do not hesitate to contact me by e-mail at Cynthia.Ow@Olympus.com or by phone at (647) 999-3203.