Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02

Company
Celltrion Usa Inc
Recall Initiated
December 28, 2021
Posted
March 2, 2022
Terminated
June 10, 2024
Recall Number
Z-0659-2022
Quantity
243 kits (6,075 tests)
Firm Location
Jersey City, NJ

Reason for Recall

The affected lot was placed under quarantine in September 2021 due to complaints of alleged false positives, but 243 kits were erroneously released to customers in December 2021.

Distribution

US distribution to CA, FL, NJ, NY, PA, SC.

Lot / Code Info

Lot COVGCCM0008

Root Cause

Process control

Action Taken

The firm initially placed a hold on product due to complaints of false positive and false negative results in September 2021. After product was erroneously distributed from this lot in December 2021, the firm notified affected customers that the lot should not have been distributed and should not be used. This email went out on or around December 28, 2021. Customers were instructed to NOT use the affected lot.