Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02

Company
Celltrion Usa Inc
Recall Initiated
December 2, 2021
Posted
March 2, 2022
Terminated
June 10, 2024
Recall Number
Z-0680-2022
Quantity
8080 kits originally distributed; 726 kits redistributed
Firm Location
Jersey City, NJ

Reason for Recall

Returned test kits were erroneously shipped to customers. The kits are intended for Research Use Only (RUO). Additional RUO test kits were shipped to customers unlikely to use them for RUO purposes, and the firm provided a letter assuring distributors that RUO tests could be used for clinical diagnosis.

Distribution

Distributed to TX.

Lot / Code Info

Lots COVGCCF0001, COVGCCF0002

Root Cause

Employee error

Action Taken

On December 2, 2021, the firm notified the affected distributor that the two lots are Research Use Only and should not be used. The distributor then notified the customer. Customer reported all product had been used.