3mL Syringe w/Spinning Spiros, Red Cap, REF: CH2003; 5mL Syringe w/Spinning Spiros, Red Cap, REF: CH2005; 10mL Syringe w/Spinning Spiros, Red Cap, REF: CH2010; 20mL Syringe w/Spinning Spiros, Red Cap, REF: CH2020; 30mL Syringe w/Spinning Spiros, Red Cap, REF: CH2030; 60mL Syringe w/Spinning Spiros, Red Cap, REF: CH2060; 30" (76 cm) Appx 3.3 mL, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CH3011; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Red Cap, Hanger, REF: CH3011-C; 40 IN (102 cm) App 4.8 ml, Admin Set, 20 Drop, 2 Spiros, Bag Hanger, REF: CH3020; 14 IN (36 cm) App 4.8 ml, Bifuse Add-On Set, Spiros w/Red Cap, Bag Spike, REF: CH3033; 5" (13 cm) Bag Spike Adapter w/Spiros w/Red Cap, Vented Cap, REF: CH3034; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, 0.2 Micron Filter, Spiros w/Red Cap, Hanger, REF: CH3128; Oncology Kit w/60" (152 cm) Appx 2.3 ml, Smallbore Ext Set w/Spiros, Anti-Siphon Valve, ChemoClave, Clamp, Rotating Luer, Double ChemoClave Syringe Transfer Set, Spiros w/Red Cap, REF: CH3606; 5" (13 cm) Appx 0.50 ml, Smallbore Trifuse Ext Set w/3 ChemoClave (Purple, Green Rings), Spiros w/Red Cap, 3 Clamps, REF: CH3625; 86" (218 cm) Appx 10.4 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, REF: CH3642; Oncology Kit w/40" (102 cm) Appx 4.7 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, Double ChemoClave Syringe Transfer Set, REF: CH3677; 5.5" (14 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 ChemoClave, Spiros w/Red Cap, 3 Clamps, REF: CH3700; 50 IN (127 cm) Appx 6.4 ml, Ext Set w/ChemoClave, Spiros w/Red Cap, Clamp, REF: CH3703; 16 IN (41 cm) Appx 1.6 ml, Ext Set, ChemoClave, Spiros w/Red Cap, REF: CH3715; Oncology Kit, Ext w/Spiros, Grad Conn, 2 Vial Spikes w/Clave, Spinning Spiros , REF: CH3929; Oncology Kit, w/Ext, MicroClave, Clave, Grad Conn, Transfer, Spiros, Vial Spike, Clave, REF: CH3932; Oncology Kit, 12 IN Ext, Spiros, Graduated Adapter, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3943; Oncology Kit w/12" Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter, ChemoClave Vented Vial Spike, 20mm, ChemoClave Vented Vial Spike, 13mm, REF: CH3944; Oncology Kit, 16 IN (41 cm) Ext Set, ChemoClave, Graduated Conn, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3946; Oncology Kit, 21" (53 cm) Appx 3.2 ml, Ext Set w/Spiros w/Red Cap, Graduated Adapter, Locking Universal Vented Vial Spike w/ChemoClave, Closed Vial Spike, 13mm w/ChemoClave, Spinning Spiros w/Red Cap, REF: CH3949; 46" (117 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, REF: CH4002
- Company
- Icu Medical, Inc.
- Recall Initiated
- August 31, 2020
- Recall Number
- Z-0704-2021
- Quantity
- 534764
- Firm Location
- Amanecer San Clemente, CA
Reason for Recall
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Lot / Code Info
Lots: 4553702, 4605473, 4763191, 4775520, 4849205, 4549847, 4548779, 4572314, 4734247, 4763917, 4849208, 4741169, 4763921, 4560231, 4750962, 4549850, 4763200, 4734249, 4794307, 4806855, 4849220, 4861176, 4877472, 4605484, 4749780, 4795448, 4876942, 4719395, 4737146, 4764825, 4800210, 4866411, 4800214, 4734262, 4794308, 4795454, 4806862, 4849239, 4849248, 4849251, 4605504, 4719410, 4734273, 4757665, 4763982, 4799740, 4861174, 4605074, 4741282, 4751117, 4809117, 4877145, 4894538, 4557534, 4733509, 4741291, 4743763, 4774120, 4733512, 4865834, 4894550, 4573105, 4803126, 4808046, 4860027, 4806400, 4720476, 4724725, 4749443, 4786358, 4802674, 4849131, 4783296, 4730109, 4742662, 4759699, 4768019, 4794497, 4877852, 4897107, 4712378, 4749445, 4799839, 4806855, 4861176, 4719395, 4800210, 4764826, 4800214, 4719927, 4726732, 4733428, 4734262, 4741183, 4756802, 4763958, 4794308, 4794310, 4795452, 4806862, 4849230, 4849233, 4849239, 4719410, 4799839, 4859282
Root Cause
Process control
Action Taken
On 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and is actively increasing the amount of available inventory. In the event specific product is unavailable, consider use of the non-spinning Spiros or the ChemoLock CSTD as alternatives. Please contact customer service for product availability. 4) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7108 (MF, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. 5) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle to obtain a response form. For further inquiries, please contact: Global Complaint Management, 1-844-654-7780 or ProductComplaintsPP@icumed.com; Customer Service, 1-866-829-9025, option 8 or, customerservice@icumed.com, (M-F, 8am-6pm CT) On 10/07/2020, the Medical Device Recall notice, with an expanded affected device scope, was sent to customers.