RecallDepth

Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000

Company
Icu Medical, Inc.
Recall Initiated
December 19, 2025
Posted
January 16, 2026
Recall Number
Z-1111-2026
Quantity
14,684,607 items
Firm Location
Amanecer San Clemente, CA

Reason for Recall

Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body

Distribution

Worldwide - US Nationwide distribution including in the states of FL, AZ, RI, TN, KS, NJ, NY, VA, SC, GA, CA, MS, TX, OH, NC, MA, WA, IL, PA, NV, NH, LA, MI, NE, ME, MN, NM, OK, MD, IA, OR, SD, WI, MO, AL, WY, IN, KY, DC, UT, AR, WV, AK, ND, CO and the countries of BE, DE, CO, ES, AU, AR, SA, SI, IT, LK, MY, AT, FI, GB, TW, MX, NL, IE, CA, GR, ZA, SE, SK, NZ, LT, CH, HR, FR, JO, PE, SG, BR, PL, RO, HU, CZ, NO, AE, EE, CN, MA, HK.

Lot / Code Info

UDI-DI: 00840619026059, 00840619006785. REF/Lot(Expiration): 011-D1000/14015455(5/1/2029), 14044313(6/1/2029), 14228823(12/1/2029), 14274294(1/1/2030); 01C-D1000/14171414(10/1/2029), 14304670(2/1/2030); 055-D1000/13772668(9/1/2028), 13778926(9/1/2028), 13791786(10/1/2028), 13794986(10/1/2028), 13797963(10/1/2028), 13801102(10/1/2028), 13806764(10/1/2028), 13812991(11/1/2028), 13822705(11/1/2028), 13828556(11/1/2028), 13833879(11/1/2028), 13838937(12/1/2028), 13867515(1/1/2029), 13875052(1/1/2029), 13876502(1/1/2029), 13876503(1/1/2029), 13882254(1/1/2029), 13882256(1/1/2029), 13887995(1/1/2029), 13887996(1/1/2029), 13894349(2/1/2029), 13894350(2/1/2029), 13903971(2/1/2029), 13903972(2/1/2029), 14045054(6/1/2029), 14045055(6/1/2029), 14049379(6/1/2029), 14049404(6/1/2029), 14063784(6/1/2029), 14063785(6/1/2029), 14071890(7/1/2029), 14071891(7/1/2029), 14071892(7/1/2029), 14087320(7/1/2029), 14090798(7/1/2029), 14110500(8/1/2029), 14131560(8/1/2029), 14144999(9/1/2029), 14145000(9/1/2029), 14150873(9/1/2029), 14150874(9/1/2029), 14150875(9/1/2029), 14170224(10/1/2029), 14170226(10/1/2029), 14179702(10/1/2029), 14179704(10/1/2029), 14179706(10/1/2029), 14208522(11/1/2029), 14208523(11/1/2029), 14212866(11/1/2029), 14212870(11/1/2029), 14216154(11/1/2029), 14216155(11/1/2029), 14224623(11/1/2029), 14248560(12/1/2029), 14248561(12/1/2029), 14274291(1/1/2030), 1427429291/1/2030), 14278257(1/1/2030), 14278258(1/1/2030), 14278259(1/1/2030), 14278260(1/1/2030), 14300970(2/1/2030), 14300972(2/1/2030), 14303936(2/1/2030); D1000/13768000(9/1/2028), 13772666(9/1/2028), 13778922(9/1/2028), 13778923(9/1/2028), 13778925(9/1/2028), 13791783(10/1/2028), 13791784(10/1/2028), 13791785(10/1/2028), 13792328(10/1/2028), 13794980(10/1/2028), 13794982(10/1/2028), 13794984(10/1/2028), 13801099(10/1/2028), 13801100(10/1/2028), 13801101(10/1/2028), 13806761(10/1/2028), 13806762(10/1/2028), 13806763(10/1/2028), 13812988(11/1/2028), 13812989(11/1/2028), 13812990(11/1/2028), 13822693(11/1/2028), 13822697(11/1/2028), 13822701(11/1/2028), 13825588(11/1/2028), 13828548(11/1/2028), 13828551(11/1/2028), 13828553(11/1/2028), 13833876(11/1/2028), 13833877(11/1/2028), 13833878(11/1/2028), 13838931(11/1/2028), 13838933(12/1/2028), 13838935(11/1/2028), 13848442(12/1/2028), 13848444(12/1/2028), 13848445(12/1/2028), 13854647(12/1/2028), 13854648(12/1/2028), 13854649(12/1/2028), 13858987(12/1/2028), 13859857(12/1/2028), 13859858(12/1/2028), 13859859(12/1/2028), 13867512(1/1/2029), 13867514(1/1/2029), 13876507(1/1/2029), 13876508(1/1/2029), 13876509(1/1/2029), 13882259(1/1/2029), 13882260(1/1/2029), 13887205(1/1/2029), 13887206(1/1/2029), 13887997(1/1/2029), 13894351(1/1/2029), 13894352(2/1/2029), 13894353(2/1/2029), 13903974(2/1/2029), 13903975(2/1/2029), 13903976(2/1/2029), 13903977(2/1/2029), 13910270(2/1/2029), 13910272(2/1/2029), 13938363(3/1/2029), 13938364(3/1/2029), 13938366(3/1/2029), 13943704(3/1/2029), 13943708(3/1/2029), 13943712(3/1/2029), 13943720(3/1/2029), 13949234(3/1/2029), 13953949(3/1/2029), 13953954(4/1/2029), 13953956(4/1/2029), 13953958(4/1/2029), 13955847(4/1/2029), 13959937(4/1/2029), 13959944(4/1/2029), 13959950(4/1/2029), 13959956(4/1/2029), 13971720(4/1/2029), 13971744(4/1/2029), 13971752(4/1/2029), 13971761(4/1/2029), 13971773(4/1/2029), 13979279(4/1/2029), 13979290(4/1/2029), 13979293(4/1/2029), 13979298(4/1/2029), 13979303(4/1/2029), 13979306(4/1/2029), 13979308(4/1/2029), 13982887(4/1/2029), 13982888(4/1/2029), 13986330(4/1/2029), 13986339(4/1/2029), 13986376(5/1/2029), 13986393(4/1/2029), 13986399(4/1/2029),13986406(4/1/2029), 13986440(4/1/2029), 13999996(5/1/2029), 13999997(5/1/2029), 13999999(5/1/2029), 14000000(5/1/2029), 14000001(5/1/2029), 14000002(5/1/2029), 14000003(5/1/2029), 14015459(5/1/2029), 14015462(5/1/2029), 14015463(5/1/2029), 14015464(5/1/2029), 14015467(5/1/2029), 14021608(5/1/2029), 14021610(5/1/2029), 14021611(5/1/2029), 14021612(5/1/2029), 14021613(5/1/2029), 14027546(5/1/2029), 14027547(5/1/2029), 14027548(5/1/2029), 14027549(5/1/2029), 14027551(5/1/2029), 14027552(5/1/2029), 14027553(5/1/2029), 14036497(5/1/2029), 14036498(5/1/2029), 14036499(5/1/2029, 14036500(6/1/2029), 14036501(6/1/2029), 14036502(6/1/2029), 14036503(6/1/2029), 14037604(6/1/2029), 14037605(6/1/2029), 14041790(6/1/2029), 14041792(6/1/2029), 14041793(6/1/2029), 14041795(6/1/2029), 14041797(6/1/2029), 14041798(6/1/2029), 14041800(6/1/2029), 14056392(6/1/2029), 14056393(6/1/2029), 14056395(6/1/2029), 14056397(6/1/2029), 14056398(6/1/2029), 14056399(6/1/2029), 14056400(6/1/2029), 14063790(6/1/2029), 14063791(7/1/2029), 14063792(7/1/2029), 14071894(7/1/2029), 14071895(7/1/2029), 14087321(7/1/2029), 14087322(7/1/2029), 14087323(7/1/2029), 14090800(7/1/2029), 14090801(8/1/2029), 14097661(8/1/2029), 14097662(8/1/2029), 14100193(8/1/2029), 14100198(8/1/2029), 14110501(8/1/2029), 14110502(8/1/2029), 14110503(8/1/2029), 14110505(8/1/2029), 14115072(8/1/2029), 14115073(8/1/2029), 14115074(8/1/2029), 14115075(8/1/2029), 14115076(8/1/2029), 14131561(8/1/2029), 14131562(9/1/2029), 14131563(9/1/2029), 14131564(9/1/2029), 14135720(9/1/2029), 14135721(9/1/2029), 14135722(9/1/2029), 14135723(9/1/2029), 14135725(9/1/2029), 14145003(9/1/2029), 14145004(9/1/2029), 14145005(9/1/2029), 14147959(9/1/2029), 14150877(9/1/2029), 14150878(10/1/2029), 14150880(9/1/2029), 14162155(10/1/2029), 14170227(10/1/2029), 14170231(10/1/2029), 14170232(10/1/2029), 14179721(10/1/2029), 14183938(10/1/2029), 14183940(10/1/2029), 14183941(10/1/2029), 14192702(10/1/2029), 14192703(10/1/2029), 14201160(11/1/2029), 14201161(11/1/2029), 14201162(11/1/2029), 14201163(11/1/2029), 14201165(11/1/2029), 14216156(11/1/2029), 14216157(11/1/2029), 14219693(11/1/2029), 14219694(11/1/2029), 14224627(11/1/2029), 14224630(11/1/2029), 14226006(11/1/2029), 14226007(11/1/2029), 14226008(11/1/2029), 14226009(11/1/2029), 14226010(11/1/2029), 14226011(12/1/2029), 14226012(12/1/2029), 14226013(12/1/2029), 14228825(12/1/2029), 14228826(12/1/2029), 14228828(12/1/2029), 14228832(12/1/2029), 14244970(12/1/2029), 14244971(12/1/2029), 1424497(12/1/2029), 14244977(12/1/2029), 14244978(12/1/2029), 14244980(12/1/2029), 14251976(12/1/2029), 14251977(12/1/2029), 14251979(12/1/2029), 14251980(12/1/2029), 14251983(12/1/2029), 14251984(12/1/2029), 14251985(1/1/2030), 14251986(12/1/2029), 14265939(1/1/2030), 14265940(1/1/2030), 14265941(1/1/2030), 14265942(1/1/2030), 14265943(1/1/2030), 14289207(1/1/2030), 14292935(2/1/2030), 14292936(2/1/2030), 14300977(2/1/2030), 14300979(2/1/2030), 14307683(2/1/2030), 14307684(2/1/2030), 14307685(2/1/2030); LAT-D1000/13768003(9/1/2028), 13778927(10/1/2028), 13812992(11/1/2028), 13848447(12/1/2028), 13867516(1/1/2029), 14015450(5/1/2029), 14044312(6/1/2029), 14063789(6/1/2029), 14090799(7/1/2029), 14145002(9/1/2029), 14228820(12/1/2029), 14228821(12/1/2029), 14228822(12/1/2029), 14274293(1/1/2030), 14274302(1/1/2030), 14307682(2/1/2030)

Root Cause

Process control

Action Taken

On 12/19/2025, correction notices were emailed to customers who were asked to do the following: 1) Destroy all affected products following your institution's process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2) Share this notification with all potential users of the device, to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there. 3) Complete and return the Customer Response Form via email to marketaction@mailac.custhelp.com For further inquiries, please contact the firm: Global Complaint Management: productcomplaintspp@icumed.com 1-(866)-216-8806; Customer Service: customerservice@icumed.com, 1-(800)-258-5361