CEI Cutting Edge Instruments (PTC-57C)
- Company
- Coltene Whaledent Inc
- Recall Initiated
- October 20, 2020
- Recall Number
- Z-0467-2021
- Quantity
- 556
- Firm Location
- Cuyahoga Falls, OH
Reason for Recall
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.
Distribution
Sold to distributors located in FL, NV, PA, TN, TX. Foreign distribution to Canada.
Lot / Code Info
J88954
Root Cause
Under Investigation by firm
Action Taken
On October 20, 2020, the firm emailed Urgent Medical Device Recall letters to affected distributors. Customers were informed that the incorrect dental carbide bur was manufactured and subsequently packaged and distributed. Customers were asked to return the product to Coltene/Whaledent, Inc. by following the instructions in the letter. The firm will replace the product at no cost. Please inform the firm if you have already used the product. Distributors were asked to either identify and notify their customers who were shipped the affected product or to provide a list of affected customers to the recalling firm so that Coltene/Whaledent, Inc. could notify them.