SciCan STATIS SL S10001 Dental Handpieces-Intended for the removal of carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations and for polishing teeth. Model Number: S10001

Company
Coltene Whaledent Inc
Recall Initiated
January 20, 2023
Posted
March 2, 2023
Recall Number
Z-1208-2023
Quantity
3896 units
Firm Location
Cuyahoga Falls, OH

Reason for Recall

Screw connection between the turbine head and the push button may loosen and fall off during the treatment, and increase the risk of swallowing or inhaling part of the turbines head or the push button

Distribution

Nationwide Foreign: Canada Germany Indonesia Ireland Lebanon Saudia Arabia Turkey

Lot / Code Info

UDI-DI: +ESANS100010 All Serial Numbers

Root Cause

Other

Action Taken

Coltene issued Urgent Medical Device Correction to Distributors. The written notification included the Field Safety Notice issued by the manufacturer and the Correction Notice that was issued on January 20th, 202 via Email. Letter states reason for notification, health risk, and action to take: Warning: When using the STATIS SL turbines with REF 2.000.7300 / S10001, with the lack of care or age-related wear and tear, there is a possibility that the push button on the turbine will come loose and this will increase the risk of swallowing or inhaling part of the turbines head or the push button. Recommended action for users: Please urgently observe the following safety instructions from the instructions for use on handling, care and running behaviour of the turbine to protect your patients. This notice needs to be passed on to all those who need to be aware, both within your own organization and in organizations where the potentially affected devices have been transferred. Please ensure that this information and the associated action are followed for a reasonable period of time to ensure the effectiveness of the action.

More recalls by Coltene Whaledent Inc

View all recalls by this company →