ParaPost Plus 25 Titanium Alloy Spiral Vented and Serrated Posts Catalog Number: P2845B Dental Posts are implanted into teeth for dental restoration purposes. The normal procedure requires a hole to be drilled into the tissue of a tooth with an intraoral drill prior to installation.

Company
Coltene Whaledent Inc
Recall Initiated
August 21, 2018
Terminated
April 10, 2020
Recall Number
Z-0376-2019
Quantity
30 packets
Firm Location
Cuyahoga Falls, OH

Reason for Recall

ParaPost Plus Titanium Posts contain stainless steel post that are a size 4

Distribution

US Distribution to states of: CA, IL, NV, NY, PA, and WA' and to country of: Canada.

Lot / Code Info

Batch I33012

Root Cause

Process control

Action Taken

The firm, Coltene, issued an "Urgent Medical Device Recall" Letter dated August 20, 2018 to the customers on August 21st via email with a delivery and read receipt requested. The letter described the product, problem and actions to be taken. The customers were instructed to return the product to Coltene/Whaledent Inc.; return any unused product and we will replace your product at no cost, and complete and return the Return Response Form via email: tricia.cregger@coltene.com or Phone: 330-916-8817. If you should have additional questions, please contact Regulatory Affairs Manager, at 330-916-8904 and Tricia.Cregger@coltene.com.