Li-ion battery packs used in BIS Vista and BIS View Monitoring Systems. Product Number/CFN: 186-0208 UDI-Device Identifier (GTIN/UPN): 10884521130319

Company
Covidien Llc
Recall Initiated
July 26, 2018
Recall Number
Z-2958-2018
Quantity
BIS 46,961; Spare Batteries: 6,937
Firm Location
North Haven, CT

Reason for Recall

Revised replacement instructions for the Li-ion battery packs used in BIS Vista and BIS View Monitoring Systems

Distribution

Nationwide Foreign: Australia, New Zealand

Lot / Code Info

Batteries with date code 3114 or older (Week 31 of 2014)

Root Cause

Device Design

Action Taken

Medtronic U.S. consignees were notified via Federal Express or certified mail on July 26, 2018, and the letter informs customers to inspect their batteries. Batteries with date code 3114 or older (Week 31 of 2014) exceed 4 years of age and should be removed from the monitor and disposed of according to facilitys standard process. Included with this notification are stickers to be placed on BIS" monitors when the battery pack is removed the stickers No Battery Installed should be placed on the monitors. The BIS Vista and View monitoring systems may be used without a battery pack only if connected to AC power source with emergency power backup system (Red Outlet).