Polysorb Braided Absorbable Suture 2/0 30 VIOLET GS-21, Item Code CL811 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in cardiovascular or neural tissue.

Company
Covidien Llc
Recall Initiated
December 18, 2019
Posted
February 19, 2020
Terminated
August 16, 2022
Recall Number
Z-1276-2020
Quantity
139,938 units total
Firm Location
North Haven, CT

Reason for Recall

There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.

Distribution

US Nationwide distribution.

Lot / Code Info

Lot - A9J0075Y Expiration Date - 2024-08-31

Root Cause

Equipment maintenance

Action Taken

On December 18, 2109, Medtronic issued Urgent Medical Device Recall and Return Form notices to customers via courier service. Customers were advised to take the following required actions: 1. Please immediately quarantine and discontinue use of the affected item code and lot listed within the notice. 2. Please return affected product as indicated within the notice. All unused products from the affected item codes and lots must be returned. 3. If you have distributed any of the Polysorb braided or Biosyn monofilament absorbable sutures listed on the recall notice, please promptly forward the information from the letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory.