GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

This recall has been terminated (originally issued July 3, 2019).

Company
Xcision Medical Systems, Llc
Recall Initiated
July 3, 2019
Terminated
July 1, 2021
Recall Number
Z-2081-2019
Quantity
2
Firm Location
Columbia, MD
Official Source
View on FDA website ↗

Reason for Recall

One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

Distribution

The products were distributed to the following US states: MD and TX.

Lot / Code Info

Serial Numbers GP0002, GP0004; UDI: (01) 00850002615005

Root Cause

Nonconforming Material/Component

Action Taken

The firm sent a Field Correction Notice letters dated 7/3/19 to customers. The letter identified the affected product, problem and actions to be taken. If you need any further information or support concerning this issue, please contact Xcsion Customer Support at 855-924-7466.