GammaPod - Treatment Planning System-a component of the GammaPod Irradiation System. Ref: XMSGP030A09.07-0.01 Intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

This recall is currently active, issued January 18, 2024. It was issued by Xcision Medical Systems, Llc.

Company
Xcision Medical Systems, Llc
Recall Initiated
November 30, 2023
Posted
January 18, 2024
Recall Number
Z-0750-2024
Quantity
4 units
Firm Location
Columbia, MD
Official Source
View on FDA website ↗

Reason for Recall

Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected

Distribution

Worldwide - US Nationwide distribution in the states of MD, PA, TX and the country of Italy.

Lot / Code Info

00850002615029(11)230321(21)TPS0018 (Distributed in the US) 00850002615029(11)230629(21)TPS0013 (Distributed in the US) 00850002615005(11)21H24(21)TPS0016 (Distributed in the US) 00850002615029(11)210629(21 )TPS0014 (Italy)

Root Cause

Software design

Action Taken

Xcision issued Urgent Field Correction Letter via email on 11/30/23. Letter states reason for recall, health risk and action to take: Please post this notification internally for awareness of GammaPod BY CUSTOMER/USER Treatment Planning System users. Xcision is working on an update to the TPS software to implement controls to identify not only the patient identification but also the series of which the structures were delineated. This change is expected for implementation in early 2024. If you need any further information or support concerning this issue, please contact Xcision Customer Support at 855-924-7466 (1-855- SUPPORT XCISION).