Cobas 8000 Modular Series system; Software Version 05-02 and 06-03; UDI: 05641446001

This recall has been terminated (originally issued July 28, 2017).

Company
Roche Diagnostics Corporation
Recall Initiated
July 28, 2017
Terminated
June 10, 2020
Recall Number
Z-0675-2018
Quantity
399
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

A software failure may incorrectly set the system settings to "default" settings, creating a risk of incorrect results.

Distribution

Distributed in 36 states: AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, WI and Puerto Rico.

Lot / Code Info

Serial numbers: 1030-08 1492-06 1492-08 15R9-05 15R9-06 15M9-04 15P2-08 16W0-09 16U1-05 16U1-09 16U1-10 16W4-07 1492-07 1167-07 1167-08 1206-05 1216-18 15F3-07 15H4-06 14A2-06 15F3-04 15F3-05 1015-05 14D2-03 14C9-06 14C9-09 1022-09 1020-01 1022-01 1022-05 1030-10 1029-09 1027-02 1362-01 1350-08 1350-07 1364-04 1364-05 1485-07 1485-08 1498-01 14A2-08 14C6-01 14D1-07 1207-09 15G6-09 15G6-10 15H4-07 1027-04 14C2-04 14C4-05 16Y5-09 16Y5-10 16Y9-07 1031-04 1138-04 15R9-01 15R9-02 1353-03 1344-09 1344-10 1364-08 1372-01 15R9-03 1366-03 1497-10 16Y6-01 16W0-08 14C6-02 15M6-07 15P9-06 15M9-03 15P9-05 16U2-01 16U2-03 16U4-09 16U4-10 16U5-01 16U5-02 16S8-10 16V9-03 16X5-02 16Y3-10 1034-10 1036-07 1138-03 1144-01 1149-01 16X5-05 1017-04 15H5-04 15H5-05 1497-04 1497-03 1142-01 1032-10 16S8-06 16W0-06 1167-10 1332-02 1332-01 1034-09 1214-19 15S3-10 16S8-03 1366-01 1366-02 1214-11 1214-16 16U5-03 16U5-04 1168-02 1168-03 1167-06 1326-17 16Y3-09 16Z2-02 14C9-10 15F3-08 15F3-09 1487-06 1485-09 1140-09 1140-10 1639-01 1492-09 14A2-05 14A2-07 1142-05 1142-06 16S8-08 16S8-09 1372-02 1497-06 1366-08 1475-09 1362-02 1362-03 16Y3-06 1497-05 1147-02 1488-01 1488-02 14D1-09 14D1-10 15H4-03 15P0-02 16S9-04 1142-04 1030-09 1142-02 1143-06 1497-08 15G7-01 15H4-05 15P9-08 15H5-01 15R8-06 15R8-07 1360-02 1360-03 1360-04 1364-06 1364-07 1497-02 16Z2-07 16Y6-04 16Y6-06 1214-20 1215-01 1019-09 1215-17 1326-02 1326-03 1326-19 1326-01 1361-06 1475-08 15H4-01 15P9-07 15R8-01 15R8-04 16U6-01 16U6-02 1350-06 1361-05 1366-06 15M6-10 15S3-08 16U6-04 1326-04 1326-13 1326-14 1326-15 1326-16 15H4-09 15H4-10 16U2-02 16X5-07 1353-02 1372-06 16Y6-02 1345-01 1345-03 1345-04 15R8-05 16U6-03 1351-02 1353-01 15H4-02 15H4-08 1207-07 1325-01 16S8-04 16S8-05 16Y4-01 1206-02 1206-03 1206-04 1206-07 1343-10 1344-07 1344-08 1344-06 15M7-01 15M7-02 16U5-10 16X4-09 16X4-10 16X5-06 1214-17 1214-18 1344-03 1344-04 15H5-02 16S9-01 16X5-08 16X5-09 1326-20 1366-05 16X5-03 1332-03 1332-06 1332-13 1332-14 1332-15 15H6-06 15S3-03 15S3-04 1638-09 16Z2-08 1168-04 1171-02 1215-18 14C6-03 15M9-07 16W0-07 1353-04 1475-10 1326-18 14D5-10 1142-08 1142-09 1167-09 1325-02 1325-03 1345-05 1362-05 14D3-07 14D4-02 16Y8-08 16U2-09 16U2-10 1149-02 16S9-08 1142-10 1143-01 15P9-01 15P9-02 1639-04 1639-02 16Z2-01 14D4-09 14D4-10 1206-01 15M6-06 1146-02 1147-03 1147-04 1161-06 14D2-02 1171-01 1207-08 16U5-09 16Y4-04 16S8-07 16U2-04 15M7-03 15M7-04 15M7-05 15M7-06 16U5-05 16Y4-02 16Y4-03 15P9-03 15P9-04 16S9-05 16Z2-03 16Z2-05 16Y3-04 16Y3-05 1497-07 14C4-06 14C2-03 16X4-07 16X4-08 1171-05 15H5-06 16Y8-10 1161-05 1164-01 1361-09 1361-10 1366-07 1497-01 15L9-09 15M6-09 1161-08 1362-04 1325-14 1325-15 1325-16 15F3-06 1639-03 1147-05 1487-07 1497-09 1147-07 1147-09 16Y3-08 16Y6-05 1344-01 1344-02 1488-07 14A1-10 14D2-01 1163-10 16Z2-09 16Z8-05 15K9-08 16Y3-07 16X5-01 16Y6-10 15P5-03 15P5-04 15P6-03 15R8-08 15R8-09 15R8-10 16S9-02 16S9-03 16U6-05 16U6-06 16U6-07 1364-02 1168-01 1171-06 15P8-09 15P8-10 1143-09 1147-01 1143-10 1332-16 1366-04 1215-20 1206-06 1344-05 1150-04 1150-05 15R9-04 16Y8-05 16Z2-06 16Y8-04 1364-03 1325-13 14D5-02 14D3-08 14D5-01 1150-06 1163-07 16Y8-09 14D6-03 14D6-01 14D6-02 16Y9-01 1639-07 1639-06 15K9-09 15L0-02

Root Cause

Device Design

Action Taken

Consignees were notified by letter (receipt required) and requested to follow instructions to determine if the software issue has occurred. If the software issue is detected the consignee contacts customer support to have system reconfigured.