RecallDepth

The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.

Company
Roche Diagnostics Corporation
Recall Initiated
June 3, 2019
Terminated
August 28, 2020
Recall Number
Z-1979-2019
Quantity
8 units
Firm Location
Indianapolis, IN

Reason for Recall

Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.

Distribution

The products were distributed to the following US states: AL, IN, and KY.

Lot / Code Info

Lot 368756

Root Cause

Material/Component Contamination

Action Taken

Urgent Medical Device Correction notification letters dated 6/3/19 were sent to customers.

More recalls by Roche Diagnostics Corporation