CADD Administration Set, 21-7339-24, Rx Only, Sterile EO.

This recall has been terminated (originally issued March 30, 2017).

Company
Smiths Medical Asd Inc.
Recall Initiated
March 17, 2017
Posted
March 30, 2017
Terminated
July 3, 2019
Recall Number
Z-1767-2017
Quantity
38,415 (38145 US; 2268 OUS)
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the product. Although the text description on the IFU correctly identifies the product to include a 0.2 micron filter, the illustration on the IFU identifies the filter size as 2.0 microns.

Distribution

US:AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WV, WY. OUS: Canada

Lot / Code Info

46X817, 46X829, 46X899, 46X828, 46X896, 46X990, 46X831, 46X897, 46X1001.

Root Cause

Packaging process control

Action Taken

Smiths Medical sent an "Urgent Medical Device Field Safety Notice" dated March 13, 2017. The letter was addressed to "Risk Manager". It provided a list of the Affected devices, Risk to Health, Instructions to Distributors & Customers. Advised consignees to notify their customers. There is no need to return the devices to Smiths Medical. Requested consignees to complete the Urgent Medical Device Field Safety Notice (FSN) Response Form and return to smithsmedical3947@stericycle.com. For questions contact Stericycle at 844-265-7407. For further questions regarding this recall , please call (844) 265-7407.