SUNRISE instrument. The INFINITE F50 instrument is a 96-well absorbance reader for the measurement of light absorbance (optical density) of liquid media. The instrument is intended to be used primarily in in-vitro diagnostic analysis of samples from the human body to obtain information on physiological and pathological states. For applications in human clinical diagnostic uses, only the Magellan Tracker software is intended for the use with the instrument. Software and instrument have been validated for measurement and for the evaluation of qualitative and quantitative Enzyme-linked Immunosorbent Assays (ELISA) according to the scheduled diagnostic parameters and instrument specifications; they are therefore for professional use in in-vitro diagnostics

This recall has been terminated (originally issued August 25, 2016).

Company
Tecan Us, Inc.
Recall Initiated
July 25, 2016
Posted
August 25, 2016
Terminated
January 23, 2017
Recall Number
Z-2661-2016
Quantity
19 units
Firm Location
Morrisville, NC
Official Source
View on FDA website ↗

Reason for Recall

The batteries have the potential to overheating, melt or char.

Distribution

Worldwide Distribution - US including AK, AL, AR, CA, FL, GA, HI, KY, MA, MD, NC, NY, PA, PR, SC, TN, TX, UT, VA. International account: Canada.

Lot / Code Info

Instrument Material #16039400, PC Serial # 2CE4220L6Z, Instrument Serial # 1201008596 and Instrument Material #30087502, PC Serial #2CE3362777, Instrument Serial #1403008116; PC Serial #2CE3461RV1, Instrument Serial #1404001455; PC Serial #2CE3461RVY, Instrument Serial #1404001456; PC Serial #2CE41503MD, Instrument Serial #1405003467; PC Serial #2CE3462FX0, Instrument Serial #1405002373; PC Serial #2CE4220BW9, Instrument Serial #1407000866; PC Serial #2CE42406XP, Instrument Serial #1409007597; PC Serial #CND4426259, Instrument Serial #1412005606; PC Serial #CND442618S, Instrument Serial #1503003902; PC Serial #CND44261R2, Instrument Serial #1504007122; PC Serial #CND44273XJ, Instrument Serial #1504004857; PC Serial #CND5219GYY, Instrument Serial #1506006286; PC Serial #CND5219GYX, Instrument Serial #1509005071; PC Serial #CND5219H21, Instrument Serial #1511009478; PC Serial #CND5219GYN, Instrument Serial #1512005969; PC Serial #CND5219H68, Instrument Serial #1603000018; PC Serial #CND5219GYJ, Instrument Serial #1606001277 and PC Serial #CND5219GZW, Instrument Serial #1606001263.

Root Cause

Nonconforming Material/Component

Action Taken

Tecan contacted all customers by Safety Announcement, via certified mail starting July 25, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Because affected batteries pose a fire and burn hazard, it is extremely important that customers check whether their battery is affected. Customers were advised to go immediately to the HP Battery Recall website at: http://www.HP.com/go/batteryprogram2016 to check if their battery is affected by this recall. Customers were also advised to complete and return the Customer Response Form at QA@tecan.com or fax it to 919-287-2961. Customers with questions should contact HP at 1-888-202-4320 or Tecan Technical Service's Helpdesk at 800-832-2687, option 1.