Freedom EVO 200 - Product Usage: It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

This recall has been terminated (originally issued April 12, 2019).

Company
Tecan Us, Inc.
Recall Initiated
April 12, 2019
Terminated
January 18, 2022
Recall Number
Z-1851-2019
Quantity
40
Firm Location
Morrisville, NC
Official Source
View on FDA website ↗

Reason for Recall

Gas springs not always replaced within the interval specified.

Distribution

Worldwide distribution. US Nationwide distribution in the states CA, CO, CT, KS, NC, NJ, OR, RI, TN, TX, and VA. Country of Canada.

Lot / Code Info

Freedom EVO 200 BASE UNIT 10641200 and 10641541 Freedom EVO 200 MCA96 BASE UNIT 30020020 Freedom EVO 200 MCA384 BASE UNIT 30032020 Freedom EVO 200-2 LiHa BASE UNIT 10641202  Serial Numbers 222, 702, 714, 506000013, 511000005, 603000004, 702000004, 703000002, 703000005, 703000006, 707001360, 710000367, 801002513, 801003121, 802003688, 805002719, 809002562, 810000017, 812002002, 906005647, 908002944, 910006632, 1004001243, 1006000274, 1006009336, 1009002730, 1011003146, 1011004081, 1011004086, 1011004089, 1011004095, 1104000151, 1108006349, 1108006724, 1108006777, 1109000886, 1109001036, 1109006938, 1111008689, and 1203005259.

Root Cause

Under Investigation by firm

Action Taken

In April 2019, Tecan US sent Field Corrective Action Freedom EVO 200 letters to their US and Canadian consignees advising them of the issue; requesting that they complete and return an acknowledgement receipt; to pass the letter to any organization where the devices may be transferred and to arrange for a Tucan US representative to replace the gas springs