RecallDepth

Calix P PEEK Lumbar System, PLIF Trial

This recall has been terminated (originally issued May 17, 2016).

Company
X Spine Systems Inc
Recall Initiated
May 17, 2016
Terminated
April 4, 2019
Recall Number
Z-0125-2017
Quantity
1504 units
Firm Location
Miamisburg, OH
Official Source
View on FDA website ↗

Reason for Recall

The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.

Distribution

Product was shipped to the following states: AZ, CA, CO, FL, GA, ID, IN, KY, LA, MD, MI, MO, NC, NV, NY, PA, PR, TX & WV. Product was also shipped to the following countries: Australia, England, Germany, Italy, Mexico, Portugal & United Kingdom.

Lot / Code Info

Size, 22mm x 10mm x 6mm, Part #:, X034-0970, Lot #:, 015857-6-1, 021425, 025070, 28221, 31896, 656802; Size, 22mm x 10mm x 7mm, Part #:, X034-0971, Lot #:, 015857-6-1, 021425, 025070, 28221, 31896, 656802; Size, 22mm x 10mm x 8mm, Part #:, X034-0972, Lot #:, 015857-7-1, 021426, 025071, 28224, 656803; Size, 22mm x 10mm x 9mm, Part #:, X034-0973, Lot #:, 015857-8-1, 021427, 025072, 28225, 28225R, 656804; Size, 22mm x 10mm x 10mm , Part #:, X034-0974, Lot #:, 015857-9-1, 021428, 025073, 28226, 656805; Size, 22mm x 10mm x 11mm, Part #:, X034-0975, Lot #:, 015857-10-1, 021429, 025074, 28227, 656806; Size, 22mm x 10mm x 12mm, Part #:, X034-0976, Lot #:, 015857-11-1, 021430, 025075, 28228, 656807; Size, 22mm x 10mm x 13mm, Part #:, X034-0977, Lot #:, 015857-12-1, 021432, 025076, 28229, 656808; Size, 22mm x 10mm x 14mm, Part #:, X034-0978, Lot #:, 015857-13-1,021433, 025077,125077, 28230, 656809; Size, 22mm x 10mm x 15mm, Part #:, X034-0979, Lot #:, 656810, 015857-13, 21434, 25078, 28231; Size, 22mm x 10mm x 16mm, Part #:, X034-0980, Lot #:, 015857-15-1, 015857-16-1, 021435, 021436, 025079, 28232, 656811; Size, 26mm x 10mm x 6mm, Part #:, X034-1050, Lot #:, 015857-15-1, 015857-16-1, 021435, 021436, 025079, 28232, 656811; Size, 26mm x 10mm x 8mm, Part #:, X034-1052, Lot #:, 015857-19-1, 021439, 025081, 10346, 26234, 28234, 656813, 656813R; Size, 26mm x 10mm x 9mm, Part #:, X034-1053, Lot #:, 10121, 11947; Size, 26mm x 10mm x 10mm, Part #:, X034-1054, Lot #:, 015857-20-1, 021440, 025082, 28235, 656814; Size, 26mm x 10mm x 11mm, Part #:, X034-1055, Lot #:, 10124, 11948; Size, 26mm x 10mm x 12mm, Part #:, X034-1056, Lot #:, 015857-21-1, 021441, 025083, 28236, 656815; Size, 26mm x 10mm x 13mm, Part #:, X034-1057, Lot #:, 10125, 11949 & Size, 26mm x 10mm x 14mm, Part #:, X034-1058, Lot #:, 021443, 015857-22-1, 015857-23-1, 021442, 021443, 025084, 28237, 656816.

Root Cause

Component design/selection

Action Taken

Consignees will be notified via email and courier and will be instructed to return the affected product. Upon receipt of the recalled trays, the trials and rasps will be removed from the tray and quarantined for evenutal destruction.

More recalls by X Spine Systems Inc

View all recalls by this company →