Xpress System Pedicle Screw Assembly, 5.5 x 55mm. An internal fixation device for spinal surgery consisting of rods, pedicle screw assemblies (screw, cap, yoke) and cross bar connector assemblies.

This recall has been terminated (originally issued April 22, 2015).

Company
X Spine Systems Inc
Recall Initiated
March 11, 2015
Posted
April 22, 2015
Terminated
October 21, 2015
Recall Number
Z-1512-2015
Quantity
14 units
Firm Location
Miamisburg, OH
Official Source
View on FDA website ↗

Reason for Recall

One lot of instruments have etched on the Yoke of the Pedicle Screw a laser marking size of 7.5 x 55mm instead of 5.5 x 55mm as described on the product label.

Distribution

US Distribution to states of: KS, NV & TX.

Lot / Code Info

Lot #: 4084-01; M/N X073-5555; UDI: *+M697X07355551/$408401U*

Root Cause

Nonconforming Material/Component

Action Taken

The firm, X-spine Systems, Inc., sent an "Urgent Medical Device Recall" letter dated March 19, 2015 to their customers. The letter describes the product, problem and actions to be taken. The customers were instructed to notify affected hospital and surgeon of the recall immediately; complete the attached Medical Device Recall Return Response Form and return it with any affected product in inventory to X-Spine Systems immediately via mail: 725 Alexandersville Road, Miamisburg OH 45342-RMA #5288; and No action should be taken with product that has been implanted. Please contact customer service at 800.903.0640 or email orders@x-spine.com with any questions.