Merge OfficePACS software. The firm name on the label is Merge Healthcare, Hartland, WI. OfficePACS is intended for the manipulation and displaying of medical images by a suitable licensed and qualified healthcare professional. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. OrthoPACS also uses such images, in conjunction with templates for prosthetic devices, for the purposes of choosing the nature and characteristics of the prosthetic device to be used when planning a potential surgical procedure.

This recall has been terminated (originally issued July 8, 2016).

Company
Merge Healthcare, Inc.
Recall Initiated
January 30, 2016
Posted
July 8, 2016
Terminated
February 3, 2017
Recall Number
Z-2144-2016
Quantity
216 sites have the affected software
Firm Location
Hartland, WI
Official Source
View on FDA website ↗

Reason for Recall

Potential data loss occurs as a result of the software archiving not working properly.

Distribution

The software was distributed nationwide to medical facilities. There was no government/military/foreign distribution.

Lot / Code Info

Software versions: 4.0, 4.1, 4.1.1, 4.1.2, and 4.2

Root Cause

Software design

Action Taken

Merge Healthcare sent an Urgent Medical Device Recall letter dated January 26, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customers. Customers were instructed to provide a copy of the notification to all users. For additional questions customers were advised to call Merge Customer Service at (87) 741-5369. For questions regarding this recall call 262-367-0700.