Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.
This recall is currently active, issued March 20, 2025. It was issued by Merge Healthcare, Inc..
- Company
- Merge Healthcare, Inc.
- Recall Initiated
- February 19, 2025
- Posted
- March 20, 2025
- Recall Number
- Z-1402-2025
- Quantity
- 253
- Firm Location
- Hartland, WI
- Official Source
- View on FDA website ↗
Reason for Recall
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Distribution
US
Lot / Code Info
Lot Code: Software versions and UDIs 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206 12.1, UDI: (01)00842000100928(10)12.1(11)210122; 12.1.1, UDI: (01)00842000100928(10)12.1.1(11)211220; 12.2 UDI: (01)00842000100065(10)12.2.0(11)220713; 12.3, UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.1 UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.2 UDI: (01)00842000101109(10)12.3.2(11)230612; 12.3.3 UDI: (01)00842000101109(10)12.3.3(11)230913; 12.4 UDI: (01)00842000101123(10)12.4.0(11)231220; 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522; 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206
Root Cause
Software design
Action Taken
Merge Healthcare issued an Urgent Field Safety Notice to it's consignees on 02/19/2025 via email. The notice explained the issue, risk to patients, provided a workaround to immediately address the issue, and recommended clinicians check measurements in the final report to ensure that they accurately reflect the patient's condition pending software update. Required Actions: Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Schedule an update to implement the configuration change on your system. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge. Please forward this notice to any individuals within your organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The firm identified a missing software version (11.0.2) and issued the corrected notice on 03/03/2025.