Aircast REF 3011 - PL RX only VF-PL Sterile Cuff Pkg., Lot# 418003, Expiration Date: 2020 -12 - 31, Made in Mexico, DJO, LLC.

Company
Djo, Llc
Recall Initiated
January 17, 2019
Recall Number
Z-1985-2019
Quantity
9 units
Firm Location
St Vista, CA

Reason for Recall

The product may not meet sterility requirements. Product labeled as sterile, however all units do not appear to meet sterility requirements.

Distribution

Distributed domestically, Oklahoma

Lot / Code Info

Lot # 418003 UDI: (01)00888912302418

Root Cause

Environmental control

Action Taken

On January 18, 2019 a Urgent: Medical Device Recall letter was sent via email to customers. The letter ask all customers to send the notification to all customers who have purchased or were issued the affected product. In addition, the letter request that the customer complete the "Return Response" form by February 15, 2019. If you have any additional questions, please contact the firm at 1-800-321-9549.