RecallDepth

Monode, Shortwave Accessory Model #: 02200002 Used in conjunction with the Intelect Shortwave Diathermy 100 and Intelect Shortwave Diathermy 400 units. For the clinical application of electromagnetic energy in the radio frequency of 27.12 megahertz to generate deep heat within body tissues.

Company
Djo, Llc
Recall Initiated
August 21, 2015
Posted
October 15, 2015
Terminated
October 24, 2016
Recall Number
Z-0121-2016
Quantity
2887
Firm Location
St Vista, CA

Reason for Recall

Overheating and burning of the wire insulation inside the applicator.

Distribution

Worldwide Distribution - US Nationwide in the states of WY, OH, WA, NC, FL, NE, GA, KS CA, MO, IN, MA, IL, OK, AZ, TX, AR, UT, NJ, WI, PA, SC, WV, NY, MI, MN, VA, AK, TN, OR, NC, NH, KY, CO, NV, ID and the countries of El Salvador, China, Ecuador, South Korea, Chile, Israel, Peru, Nicaragua, Bolivia, Indonesia

Lot / Code Info

Model #: 02200002  All units received from the manufacture prior to the week of July 12, 2015

Root Cause

Device Design

Action Taken

The firm sent out two customer notification letters; one to distributors and one to direct users. Letters were sent September 1, 2015 via USPS certified return receipt. Distributors were asked to obtain a list of their customers with contact information that have purchase or received the monode accessory or Intelect Shortwave device that included the monode as an accessory item. Once the list has been assembled the firm requests that it be forwarded to DJO Global Chattanooga Product Support by phone 877-733-4888, or by e-mail at monode.support@djoglobal.com. The firm's notification states that customers will be sent an upgraded monode accessory along with a pre-paid shipper for the return of the current used unit. Any questions the letter states to contact Product Support per the contact information above or William Fisher, Global Regulatory Compliance at 760-734-3126. Letters sent to direct users received the same instruction of returning the product for a replacement. Any questions the letter states to call 877-733-4888 or email at monode.support@djoglobal.com.