RecallDepth

3M Coban Self-Adherent Wrap Shipper carton label identified the product as 3M Coban LF Latex Free Catalog #2081, lot number 2018-10AM.

Company
3M Company Health Care Business 3M Center
Recall Initiated
December 1, 2015
Terminated
May 24, 2016
Recall Number
Z-0601-2016
Quantity
167 cases (30 rolls/case)
Firm Location
Saint Paul, MN

Reason for Recall

Product mislabeled as latex free. Actual Catalog number 1581, lot 2020-10AN. contains latex. This could pose a health risk to users with sensitivity or allergic reaction to latex, which could range from mild to severe.

Distribution

US nationwide distribution.

Lot / Code Info

Catalog# 1581, lot # 2020-10AN Catalog# 2081, lot # 2018-AM (shipper carton)

Root Cause

Process control

Action Taken

Consignees were sent via certified mail on 12/1/2015 a 3M letter dated November 25, 2015 and an emal on 12/5/2015 for the updated 3M "Urgent Medical Device" letter dated November 25, 2015. The letter was addressed to 3M Health Care Distributor. The letter described the problem, device involved in the recall and the action required. Advised consignees to quarantine the product, notify their customers and to complete and return the "Product Recall Form". Distributors were also provided with a "Customer" letter for distribution. For questions contact 3M Customer Helpline 1-800-228-3957.

More recalls by 3M Company Health Care Business 3M Center

View all recalls by this company →