RecallDepth

Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837

Company
3M Company Health Care Business 3M Center
Recall Initiated
November 22, 2024
Posted
December 23, 2024
Recall Number
Z-0743-2025
Quantity
171 orders
Firm Location
Saint Paul, MN

Reason for Recall

Attachments may not mate with the tooth as intended in the digital treatment design.

Distribution

Product was distributed nationwide US, and internationally to Canada, Australia, and New Zealand.

Lot / Code Info

Only 3M Clarity Precision Grip Attachment Dual Arch Attachments are potentially affected. Part numbers/UDI-DI: 359-830/ 00652221307286; 359-831/ 00652221307293; 359-832/00652221307309; 359-833/ 00652221307316; 359-834/ 00652221307361; 359-835/ 00652221307378; 359-836/ 00652221307385; 359-837/ 00652221307392

Root Cause

Software design (manufacturing process)

Action Taken

On November 22, 2024, the firm notified customers via URGENT Product Recall letters. Customers were informed that select grip attachments were impacted by a software nonconformance, limited to the lower arch of specific dual-arch Clarity Precision Grip Attachment orders. The firm advises that customers not use products from the identified Case IDs, and that they should dispose of affected products. Customers should complete and return the included customer product destruction form. The firm will provide replacement product. If you have additional questions or need to report an adverse event, contact the Clinician Support team (800-423-4588, option 2) or customortho@solventum.com.