NxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5

Company
Nxstage Medical, Inc.
Recall Initiated
October 29, 2015
Terminated
April 29, 2019
Recall Number
Z-0326-2016
Quantity
323 units
Firm Location
Lawrence, MA

Reason for Recall

Ultrafiltration (UF) Volume software error inaccurate fluid removal

Distribution

Worldwide Distribution. US Nationwide, Netherlands, and Sweden.

Lot / Code Info

Software versions 4.9 and 4.10

Root Cause

Software Design Change

Action Taken

NxStage issued issued a Field Safety Notice (FSN) on 10/29/15 to consignees who have NX1000-5 or a NX1000-5-A cycler. The FSN included Instructions for steps to be taken so that the cycler and treatments are not affected by the software error. NxStage will be sending a representative to the facility to attach detailed instructions on each cycler as a user reference. Return the attached reply form to acknowledge receipt of this letter. Questions or comments, please contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243).