NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
- Company
- Nxstage Medical, Inc.
- Recall Initiated
- April 28, 2025
- Posted
- May 28, 2025
- Recall Number
- Z-1864-2025
- Quantity
- 115,465 cases (230,930 units)
- Firm Location
- Lawrence, MA
Reason for Recall
Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
Distribution
US Nationwide distribution.
Lot / Code Info
a. UDI-DI number: M535RFP4000. Device Catalog Number RFP-400. Lot Numbers: Q2407307, Q2407308, Q2407309. b. UDI-DI number: M535RFP4010. Device Catalog Number RFP-401. Lot Numbers: Q2407061, Q2407147, Q2407148, Q2407149, Q2407313, Q2407386, Q2407387, Q2407388, Q2407389, Q2407390. c. UDI-DI number: M535RFP4020. Device Catalog Number RFP-402. Lot Numbers: Q2407213, Q2407214. d. UDI-DI number: M535RFP4030. Device Catalog Number RFP-403. Lot Number: Q2407311. e. UDI-DI number: M535RFP4040. Device Catalog Number RFP-404. Lot Numbers: Q2407062, Q2407063, Q2407064, Q2407065, Q2407067, Q2407068, Q2407069, Q2407070, Q2407071, Q2407217, Q2407218, Q2407219, Q2407220, Q2407221, Q2407222, Q2407223, Q2407224, Q2407225, Q2407301, Q2407302, Q2407303, Q2407304, Q2407305, Q2407306, Q2407391, Q2407392, Q2407405, Q2407406. f. UDI-DI number: M535RFP4060. Device Catalog Number RFP-406. Lot Numbers: Q2407072, Q2407137, Q2407140, Q2407141, Q2407216, Q2407310, Q2407312. g. UDI-DI number: M535RFP4530. Device Catalog Number RFP-453. Lot Number: Q2407066. h. UDI-DI number: M535RFP4540. Device Catalog Number RFP-454. Lot Numbers: Q2407145, Q2407146, Q2407226. j. UDI-DI number: M535RFP4560. Device Catalog Number RFP-456. Lot Numbers: Q2407138, Q2407139, Q2407142, Q2407143, Q2407144, Q2407227.
Root Cause
Under Investigation by firm
Action Taken
NxStage notified consignees via mail on 04/28/2025. Consignees were instructed to check inventory for affected units and cease use if any are on hand, complete and return the response form, and contact NxStage Customer Service to arrange return of all affected units.