Toshiba Aquilion CT System TSX-101A

This recall has been terminated (originally issued October 9, 2015).

Company
Toshiba American Medical Systems Inc
Recall Initiated
May 8, 2015
Posted
October 9, 2015
Terminated
October 26, 2016
Recall Number
Z-0013-2016
Quantity
459
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved.

Distribution

US Distribution including Puerto Rico

Lot / Code Info

system software V6.0 or later

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Toshiba America Medical Planned Action to bring defect into Compliance: a. A notification will be sent to all consignees with affected systems in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on each system to prevent this occurrence from happening. Each consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. Each consignee will be advised to do the following until the corrective measures have been taken, i.e., revised software installed on their systems. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Toshiba America Medical Systems, Inc. will implement this CAP by February 1, 2016. If you have any questions please contact your local Toshiba Representative at (800) 521-1968.

More recalls by Toshiba American Medical Systems Inc

View all recalls by this company →