Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.

This recall has been terminated (originally issued January 4, 2018).

Company
Toshiba American Medical Systems Inc
Recall Initiated
January 4, 2018
Terminated
October 10, 2023
Recall Number
Z-2315-2018
Quantity
162 units total
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.

Distribution

USA (nationwide) Distribution to the states of : CA, CO, DE, GA, HI, IA, ID, IL, IN, MD, MI, MN, MO, NC, ND, NV, OH, OR, PA, SC, SD, TN, TX, VA, WA, WV and WY.

Lot / Code Info

Model No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1732013, HAB1732016, HAB16Z2002, NJC17X2086, NJC17X2087, JHA16X2004, HAB1732027, HAB16Z2001, HAB16Z2002, NJB1762003, JHA1752022, HAB16Z2003, HAB16Z2004, JHA1752019, NJB1782050, HAB1742033, NJB1772031, HNA1792013, HNA1772008, JHA1752010, HAB1742035, HAB1742036, NJB1762001, NJB1782042, JHA1762034, HAB1732030, HAB16Z2005, HAB1732007, HAB1742034, HAB1742037, JHA1752024, NJB1772019, JHA1762029, JHA1762031, NJB1782061, HAB1732019, HNA1772001, HAB1712008, HAB1722011, HAB1712010, HAB1712009, HNA1792014, HAB1722012, HAB16Z2003, HAB1732017, HAB1732018, HAB1712006, HAB1732014, HAB1712007, HAB1732015, HAB1732021, HAB1732020, HAB1732025, JHA1762030, HAB1732029, HAB1732031, NJC17X2066, HNA1772002, NJB1782060, NJB1782043, NJB1762002, JHA16X2002, NJB1782057, NJB1782055, JHA1752021, NJB1782058, NJB1762006, NJB1782056, NJC17X2089, NJB1772023, NJB1782051, NJB1782052, NJB1782053, NJB1782054, HAB1732006, NJB1782062, NJB1782063, NJB1762004, NJB1762005, NJB1772022, JHA1752014, JHA1752016, JHA1752017, JHA1752018, HNA1792012, NJC17X2068, NJC17X2069, NJC17X2070, NJC17X2067, NJB1782046, NJB1782047, NJB1782048, NJB1782049, NJB1772039, NJB1782041, NJB1782045, NJC17X2083, NJC17X2084, NJC17X2081, NJC17X2079, NJC17X2080, NJC17X2082, NJC17X2076, NJC17X2077, NJC17X2078, NJB1772021, NJB1782044, JHA1752020, HNA1772005, NJB1782064, JHA1762032, HAB1732010, HAB1732008, NJB1782059, NJB1762008, NJB1762007, NJB1762009, NJB1772024, NJB1772014, NJB1772017, NJB1772018, NJB1772015, NJB1772016, NJB1772011, NJB1772012, NJB1772020, NJB1762010, NJB1772035, NJB1772036, NJB1772037, NJB1772038, NJB1772026, NJB1772027, NJB1772028, NJB1772030, NJB1772034, NJB1772025, NJB1772032, NJB1772013, JHA1762036, NJC17X2090, JHA1762033, HNA1772010, HNA1772009, NJB1772029, HNA1772004, HNA1772003, NJB1782065, NJC17X2071, NJB17822040, NJC17X2074, NJC17X2075, NJC17X2085, HAB1732026, JHA16X2007, HAB1732022, JHA1752023, JHA1762026, JHA1762025, HAB1732028, JHA16X2001, JHA16X2003, HAB1742011, HAB1742012, JHA16X2001, JHA16X2003, JHA1752004, JHA1752005, JHA1752006, JHA1752007, JHA1752008, JHA1752009, JHA1752010, JHA1752011, JHA1752012, JHA1752013, NJB1772033, NJC17X2072, NJC17X2073, NJC17X2088, HAB16Z2001, HAB16Z2004, HAB16Z2005, HAB1732024, HAB1732032, HNA1792015, HNB17X2016, HNB17X2017, HNB17X2018, HNB17X2019, HNB17X2020, HNB17X2021, HNB17X2022, HNB17X2023, HNB17X2024, HNB17X2025, HNB17X2026, HNB17X2027, HNB17X2028, HNB17X2029, HNB17X2030, JHA16X2002, JHA16X2005, JHA16X2006, JHA16X2008, JHA16X2009, JHA1752011, JHA1752012, JHA1752013, JHA1752015, JHA1762027, JHA1762028, JHA1762035, JHA1762037, JHA1762038, JHA1762039, HAB1742001, HAB1742002, JHA16X2001

Root Cause

Software design

Action Taken

Canon sent an Urgent - Medical Device Correction letter dated January 4, 2018. Customers were notified of the issue and informed that a modified firmware will be installed to correct the issue. When that modified firmware becomes available, the Toshiba service representative will contact customers to schedule an appointment to install the firmware. Until the modified firmware is installed, instructions were provided to confirm that the main power switch is sent to OFF, insert the power plug into the power outlet, and to set the main power switch to ON. The system should be operated according to the procedure described in section 9.1, Connection the Power Cable and Protective Grounding, in the operation manual. Additionally, customers were asked to complete and return the attached Customer Reply Form. For further questions, please call (714) 730-5000.