RecallDepth

JustRight 5mm Reload, JR-REL25-2.0 Reload is packaged in cases of 12., placed in a tray and sealed with a Tyvek lid. Product Usage: The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatric, and thoracic surgery for the resection, transection, and creation of anastomosis.

This recall has been terminated (originally issued July 2, 2015).

Company
Justright Surgical, Llc
Recall Initiated
June 11, 2015
Posted
July 2, 2015
Terminated
August 26, 2015
Recall Number
Z-1956-2015
Quantity
1452
Firm Location
Boulder, CO
Official Source
View on FDA website ↗

Reason for Recall

JustRight Surgical is recalling various lots of JustRight 5mm Stapler product because they may deliver incomplete staple formation and incomplete tissue closure.

Distribution

Distribution - US Nationwide and Ireland.

Lot / Code Info

Model Number JR-REL25-2.0 Lot numbers 75ED0114, 75JD1403, 75KD0304, 75KD0726, 75LD1224, 75AE0707, 75AE1406, 75BE1310, 75BE1313, 75CE2605, 75DE0802, and 75DE1524.

Root Cause

Process design

Action Taken

JustRight Surgical sent an Urgent Medical Device Recall letter dated June 12, 2015 to aaffected customers and distributor via an email. The letter identified the affected product, problem and actions to be taken. Custers were instructed Review, complete, and return the enclosed Acknowledgement Form directly to Rep, your Sales Representative or email to orders@justrightsurgical.com or fax to (720) 287-7135. The following assistance is available: " For questions related to the issue or product replacement, please contact your local sales representative, Rep, or contact JustRight Surgical Customer Service at (866) 683-1743, Monday through Friday from 7:30 AM to 5:00 PM MST. " For questions related to product pick-up by JustRight Surgical, please also contact your local sales representative or Customer Service (866) 683-1743. Return instructions are provided on the Acknowledgement Form. " For questions related to the Acknowledgement Form and its return, please contact the JustRight Surgical Director of Quality Systems at (720) 287-7146.